36
SV 200A User Manual
5
Confirm obtained calibration factor by clicking
Yes
. New calibration factor will replace the previous
one in the instrument’s memory and will be applied to all subsequent measurement results.
4.5 S
YSTEM CHECK WITH THE USE OF ELECTROSTATIC ACTUATOR
The electrostatic actuator
is used for remote system check of the instrument, which enables checking
the acoustical measurement input of the instrument.
The System Check procedure consists of a sequence of measurements of background noise and a level
generated by the electrostatic actuator (94dB). The measurement of background noise is carried out
before and after the measurement of a signal level from the actuator. It is assumed that a one-second
RMS(C) of the background noise measured for 3 consecutive seconds must be at least 20 dB lower
than the nominal level generated by the electrostatic actuator (94dB). If this condition is not met, system
check is unsuccessful. If the background condition is met, the station switches on the actuator and waits
for stable one-second RMS(C) values with an accuracy of
±0.1dB. The stabilized RMS(C) value of the
signal from the actuator cannot deviate from the nominal value (94dB) more than
±1dB. The system
check result (OK or Failed) together with the measured levels the background noise and the signal from
the actuator is recorded in the calibration and system check history file. The duration of the system
check sequence is typically <15s and can be extended up to <25s in the case of longer stabilization
period of the RMS(C) generated by the actuator. In the case of the automatic system check function,
when the test result is negative, it is repeated 4 times more every 1 minute to obtain a positive test
result. If the result is still negative, subsequent attempts are abandoned to the next planned system
check.
The measurements are paused for the duration of the system check procedure (active pause).
Terminate button
Calibration factor
Содержание SV 200A
Страница 123: ...123 SV 200A User Manual Appendix A REMOTE CONTROL firmware revision 1 x x Under development ...
Страница 153: ...153 SV 200A User Manual Free Field Frequency response of SV 200A ...
Страница 164: ...164 SV 200A User Manual 250 Hz 315 Hz 90 deg 0 deg 90 deg 0 deg 400 Hz 500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 165: ...165 SV 200A User Manual 630 Hz 800 Hz 90 deg 0 deg 90 deg 0 deg 1000 Hz 1250 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 166: ...166 SV 200A User Manual 1600 Hz 2000 Hz 90 deg 0 deg 90 deg 0 deg 2240 Hz 2500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 167: ...167 SV 200A User Manual 2800 Hz 3150 Hz 90 deg 0 deg 90 deg 0 deg 3550 Hz 4000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 168: ...168 SV 200A User Manual 4500 Hz 5000 Hz 90 deg 0 deg 90 deg 0 deg 5600 Hz 6300 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 169: ...169 SV 200A User Manual 7100 Hz 8000 Hz 90 deg 0 deg 90 deg 0 deg 8500 Hz 9000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 170: ...170 SV 200A User Manual 9500 Hz 10000 Hz 90 deg 0 deg 90 deg 0 deg 10600 Hz 11200 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 177: ...177 SV 200A User Manual 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 178: ...178 SV 200A User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 179: ...179 SV 200A User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...