SV 104IS USER’S MANUAL______________________________________________________
71
6.6
Firmware update
SVANTEK
is committed to continuous innovation path of development, and as such reserves the right
to provide firmware enhancements based on user’s feedback.
To update the instrument firmware:
Unpack the provided firmware package (provided as a suitable compressed file).
Make sure the unit is turned off and beyond docking station.
Hold down the
<PROFILE>
key and press the
<ENTER>
key to turn on the unit
.
This ensures the unit will switch on and enter the special reprogramming
BOOTLOADER
mode.
Figure 6-1
Bootloader update mode view
Place the dosimeter on the powered docking station connected via mini USB cable to the PC.
The
<USB>
text will now appear on the instrument display.
Run the
start.bat
file.
Successful firmware update will be indicated by relevant message.
Turn off the unit.
Notice:
With use of SUPERVISOR software it is very easy to check if there
are any new firmware releases available for download (see Figure 5-3 commands).
6.7
Storing the instrument
To preserve the life of the internal battery, it is recommended that the
SV 104IS
instrument
is turned off when it is stored.
It is best not to store the dosimeter for any period of time in low battery state condition.
Generally the
SV 104IS
instrument should be stored at room temperature, charged to about
40 to 60% of capacity.
After use, it is best to charge
SV 104IS
instrument in docking station charger prior to storage
for a prolonged period that is longer than 3 months.
If dosimeter is to be stored for an even longer period, it is best to take the instrument out of
storage and re-charge the dosimeter every 8 months. When the
SV 104IS
is turned off, it still
draws a small amount of battery power, so periodic recharge is to prevent battery over-
discharge.
6.8
Transportation and carrying
For transportation or storage purpose, always use the packaging provided by the manufacturer.
In a potentially dirty industrial environment it is advisable to use the carrying case provided by
the manufacturer
SA 144
, which ensures excellent mechanical and environmental protection and long term
storage conditions. The temperature range over which the device can be stored/transported is -
20°C to
+5
0°C.