SV 104A USER'S MANUAL_______________________________________________________
63
6.5.7.3
Keyboard security settings panel
The security setting enables you to protect access to the instrument when in use with a simple keypad
password to prevent users from inadvertently terminating a measurement run. This feature is set through
the
Supervisor
’s
“General”
settings tab.
To turn on the security option: switch on
“Lock During Measurement”
button. Once turned on, the
SV
104A
will disable the keyboard every time, the measurement run is started.
See the chapter 4.12 how to lock, and unlock the
SV 104A
instrument.
If
“Unlock on Key”
is set to
On
,
SV 104A
will require special code to be input by pressing four keys
defined in this panel in a particular sequence
If
“Unlock on Key”
is set to
Off
,
SV 104A
can be locked/unlocked without providing Lock/Unlock
sequence. Simply hold the
<SCROLL>
key for a couple of seconds during which a countdown
(
“Keyboard lock/unlock” 3… 2… 1…
) is displayed and the unit
gives you time to decide if you really want
to activate/deactivate the security lock. If you release the key too early, the operation will be cancelled.
Figure 6-16
Keyboard security settings panel
6.5.7.4
Auxiliary settings panel and
Bluetooth PIN
code
Under auxiliary settings panel it is possible to:
•
Enable additional warning screens to be displayed under certain conditions:
o
“Logging disabled”
it warns the operator that time history results will not be stored.
o
“Power off”
additional confirmation just before switching off (see chapter 4.3)
•
“Comment file name”
defines the name of the Voice note comment file
. “
Comment Text”
is also able to be entered here if required
•
“Vibration Marker”
is used on the basis of acceleration threshold setting (
Off, 1g,…15g
) the
lower the threshold level selected then the more sensitive the dosimeter will be to possible
bumps and shocks during use
•
“Language”
enables the user to change the menu language displayed on the screen of the
dosimeter. The default language is English
•
“Time To Automatic Shutdown”
enables the user to change time period after which the unit
will be shut down if no key will be pressed
•
“Bluetooth”
enables the user to switch on or off the Bluetooth
®
connection.
•
Note:
“Bluetooth PIN” code
configuration is required to establish connection
;
Note: For Air-transport Bluetooth
®
interface should be turned off
. Make
sure that correct settings file is applied.