36
- Buffer Size : 2048
- Hanning Window
- Exponential Averaging : Time constant is same to that of detector.
- Frequency Range : 1~100Hz
- Nth Octave : 1/1, 1/3, 1/6 Octave is useful. This parameter is changeable in [Nth Octave] of FFT
Analysis(or Spectrogram, Octave Analysis) Option.
If you want to get the more information of octave analysis, refer to the chapter V-3.
8. Setting options for Human Vibration Measurement
8.1. Selecting type of sensor, Entering sensitivity and Calibration
1
Select [Accelerometer] in [File>>Transducer Calibration…] tap.
2
Entering Sensitivity at the option [Sensitivity].
3
With connecting sensor to calibrator, click [Calibrate] and [Start] button to accomplish calibration.
>
[Compens. Gain] value is changed from 1 to another value.
4
Adjustment of [ICP Amp Gain] just before measuring : With connecting sensor and adhering the
sensor at measurement position, select [ICP Amp Gain] or [AutoRange] button to set the gain
automatically.
[Fig. IV-8-1] Selecting Sensor and Entering
Sensitivity
[Fig. IV-8-2] Calibration
8.2. Setting the measurement time and file name
When selecting [File>>Recording Option…] tap, a window appears like [Fig. IV-8-3].
1
[Save Wave File in VLM / SLM] option should be checked to store wave files.
2
[Use Timer] option should be checked to use timer. Measurement for 1 minute is recommended at
ISO6954:2000(E). If main component is smaller than 2Hz, it is recommended to set the measurement
for 2 minutes.
3
Enter the file name as you wish at [File Name].
4
Select the folder at [Select Folder:]
Содержание SLA-PA201
Страница 64: ...64 Appendix 1 The Theory for calculating RMS by ISO2631 and ISO8041 ...
Страница 67: ...67 Appendix 2 Frequency Response of Filter in the Vibration Level Meter ...
Страница 75: ...75 Appendix 3 Theory of WBCombined Filter and Frequency Response ISO6954 Filter ...
Страница 79: ...79 Fig 2 Lowpass Filter with 100Hz Cutoff Fig 3 a v transition Frequency Response ...
Страница 88: ...88 Fig 12 Error between theoretical value and measured value ...