26
2. Setting of measuring option
2. 1 Selection of Filter
2. 1. 1
Filter selection of vibration level meter
A window is opened to press
button like [Fig. IV-2-1]. You can see 9 type filters.
Here are the list below :
- 1Hz Highpass Filter
- 5Hz Highpass Filter
- 10Hz Highpass Filter
- 1KHz Highpass Filter
- 2Hz ~ 300Hz Bandpass Filter
- 2Hz ~ 1kHz Bandpass Filter
- 10Hz ~ 500Hz Bandpass Filter
- 10Hz ~ 1kHz Bandpass Filter
- WBCombined Filter (ISO6954)
[Fig. IV-2-1] Parameter Selection Window in Vibration Level Meter
And another option is supplied for low frequency analysis as below :
- BW 1.6kHz : it means 1.6kHz bandwidth for analysis, Sampling Frequency is 4096Hz.
- BW 200Hz : it means 200Hz bandwidth for analysis, Sampling Frequency is 512Hz.
- BW 100Hz : it means 100Hz bandwidth for analysis, Sampling Frequency is 256Hz.
Highpass Filter uses 2-pole Butterworth Filter, Bandpass filter uses 4-pole Butterworth Filter. And
Whole-Body Combined filter is made with being based on the ISO6954 and ISO8041.
The frequency response of the each filter is shown in the [Appendix 2 : The frequency response of
Vibration Level meter] and [Appendix 3 : The frequency response of whole-body vibration filter].
And if you select WBCombined filter, Quantity value become initialize to RMS value.
Содержание SLA-PA201
Страница 64: ...64 Appendix 1 The Theory for calculating RMS by ISO2631 and ISO8041 ...
Страница 67: ...67 Appendix 2 Frequency Response of Filter in the Vibration Level Meter ...
Страница 75: ...75 Appendix 3 Theory of WBCombined Filter and Frequency Response ISO6954 Filter ...
Страница 79: ...79 Fig 2 Lowpass Filter with 100Hz Cutoff Fig 3 a v transition Frequency Response ...
Страница 88: ...88 Fig 12 Error between theoretical value and measured value ...