LAMBDA VF-5 OPERATION MANUAL – REV. 1.04A (20190606)
43
7.
OPERATIONS: EXTERNAL CONTROL: USB INTERFACE
The Lambda VF-5 can communicate with the computer via the Universal Serial Bus (USB)
port instead of the SERIAL (RS-232) port whenever it is connected to the remote computer
with a USB cable and the appropriate USB device drivers are installed. The remote computer
must be equipped with the necessary USB hardware and its operating system must be
properly configured to recognize and work with the USB interface. Of all the USB ports
available on the remote computer, be they part of the computer itself or part of a USB hub,
one port must be available for use with the Lambda VF-5.
The Lambda VF-5 microprocessor sets the SERIAL port, by default, as the input source upon
startup and any time the ON LINE key is pressed. An appropriate command from the remote
computer through the SERIAL port will automatically set that port as the active input source
and the display on the controller will reflect that change on its bottom line. The SERIAL port
remains active until an appropriate command is sent from a remote computer through the
USB port or until the LOCAL key is pressed on the controller keypad.
There are some basic differences in the physical connections and the modes by which USB
and serial data are transmitted to the controller. The command code structures, however, are
identical. The connection and command structure required to control the Lambda VF-5 via
the USB port are discussed in this chapter.
7.1
Installing the Lambda VF-5 as a USB Device on a Windows System
The Lambda VF-5 can be used as a USB device with a remote computer that is installed with
the necessary USB hardware and is running one of the following versions of Microsoft
Windows.
1.
Windows 98
and Windows 98 SE (Second Edition)
2.
Windows ME (Millennium Edition)
3.
Windows 2000 (Professional, Server, and Advanced Server)
4.
Windows XP (Professional Edition and Home Edition)
5.
Windows Advanced Server 2003
6.
Windows Vista (all editions)
7.
Windows 7 (all editions)
7.1.1
Installation Steps
To install the Lambda VF-5 as a USB device on a remote computer running one of the above-
listed versions of the Microsoft Windows operating system, follow these steps.
1
Windows 95 and USB:
With the exception of the last releases of Windows 95 prior to the release of Windows 98, Windows 95 does NOT
support USB hardware. Those releases of Windows 95 that do support USB will generally have “With USB Support” as part of the operating
system’s title. However, the Lambda 10 series of controllers with USB interfaces have not been tested on such systems.
2
32-bit and 64-bit Versions of Windows:
The USB device driver for the Lambda SC is designed to work with 32-bit versions of the
Windows operating systems listed above. This driver is not intended for use with any 64-bit version of Windows.
3
Windows Vista:
The current USB device driver for Lambda SC has been tested under Windows Vista (32-bit version only). Because
Windows Vista’s security architecture has been significantly enhanced over previous versions of Windows, you will find that the installation
process (described next) will frequently prompt you for permission to proceed to subsequent steps.
Содержание Lambda VF-5
Страница 3: ......
Страница 4: ......
Страница 20: ...LAMBDA VF 5 OPERATION MANUAL REV 1 04A 20190606 10...
Страница 28: ...LAMBDA VF 5 OPERATION MANUAL REV 1 04A 20190606 18 This page intentionally blank...
Страница 32: ...LAMBDA VF 5 OPERATION MANUAL REV 1 04A 20190606 22 This page intentionally blank...
Страница 62: ...LAMBDA VF 5 OPERATION MANUAL REV 1 04A 20190606 52 This page intentionally blank...
Страница 64: ...LAMBDA VF 5 OPERATION MANUAL REV 1 04A 20190606 54 This page intentionally blank...
Страница 66: ...LAMBDA VF 5 OPERATION MANUAL REV 1 04A 20190606 56 This page intentionally blank...
Страница 68: ...LAMBDA VF 5 OPERATION MANUAL REV 1 04A 20190606 58 This page intentionally blank...