6
1.4.2.12
Swing Gate
The swing gate on the front of the mechanical allows the micropipette to be swung out
toward the front of the microscope for better access when changing micropipettes. The gate
can be released by loosening the fixing screw. The screw should be tightened fully during use.
Figure 1-5. Manipulator Mechanical.
1.4.3
Additional Features and Functions
1.4.3.1
‘Setup’
When the micropipette is aligned at the beginning of an experiment, the tip should be located
in the center of the microscope’s field of view and there should be as much movement range
as possible in both directions in each axis, X, Y and Z. The ‘Setup’ function has been designed
to easily achieve this and thus protect the user from inadvertently running out of motor
travel during an experiment. By following the instructions included in this manual, use of
the ‘Setup’ function will position each of the three axis motors in an optimal position to begin
working.
CAUTION: Use the Setup function only when there is no pipette and holder in the
manipulator.
The ‘Setup’ function should only be activated when there is sufficient space available around
the mechanical. Activating the ‘Setup’ function when the manipulator mechanical is too close
to a microscope component such as the condenser, may result in damage to the motor or the
microscope. Refer to Installation Procedure for details on the use of the ‘Setup’ function
during installation and Operating Instructions for details on routine use.
XENOWORKS MICROMANIPULATOR SYSTEM (TOUCHDECLUTCH) OPERATION MANUAL – REV. 4.10 20141208
(
)