8.

 

Locate the shipping screw holes. There are four through-holes in each axis slide, but only 
two of the holes will line up with the correct tapped holes on the inner slide surface when 

the instrument is in the ‘Setup’ position. 

9.

 

Locate the original shipping screws that should have been retained with the original 
packaging, and insert them into the appropriate locations. Do not over tighten the screws 

- turn them until they are lightly seated in their holes. 

10.

 

Place the mechanical in a plastic bag. 

11.

 

The mechanical is now ready for transport in the original packaging. 
Take care to ensure that the shipping screws are removed before the instrument is 

operated again (see Section 3 for the correct installation procedure). 

 
 

Содержание XenoWorks BRM

Страница 1: ...orized Micromanipulator System TouchDeclutch Operation Manual Rev 4 10 20141208 Sutter Instrument Company One Digital Drive Novato CA 94949 Voice 415 883 0128 Web www sutter com Fax 415 883 0572 Email...

Страница 2: ...RKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV ii Copyright 2015 Sutter Instrument Company All Rights Reserved XenoWorks is a registered trademark of Sutter Instrument Company 4 10 201...

Страница 3: ......

Страница 4: ......

Страница 5: ...designed for connection to a standard laboratory power outlet Overvoltage Category II and because it is a microprocessor controlled device it should be accorded the same system wiring precautions as...

Страница 6: ...nnecting the mechanical units Failure to do this may result in damage to the electronics Operate this instrument only according to the instructions included in this manual Do not operate if there is a...

Страница 7: ...rument s instrument is not recommended and if undertaken is done so at the users own risk When transporting the mechanical manipulator be sure to install the shipping screws supplied in their correct...

Страница 8: ...4 10 20141208 vi This page intentionally left blank XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV...

Страница 9: ...nsion Ring 4 1 4 2 7 Work Position 5 1 4 2 8 Home Position 5 1 4 2 9 Speed 5 1 4 2 10 Z floor 5 1 4 2 11 Micropipette Holder Clamp 5 1 4 2 12 Swing Gate 6 1 4 3 Additional Features and Functions 6 1 4...

Страница 10: ...ator 2 Figure 1 2 Joystick 3 Figure 1 3 The rear panel of the joystick 3 Figure 1 4 Joystick height adjustment 4 Figure 1 5 Manipulator Mechanical 6 Figure 1 6 Close up of axis polarity switches 7 Fig...

Страница 11: ...and ratings 11 Table C 1 Mains fuse type and ratings 31 Table D 1 Speed settings and ranges of movement 33 Table D 2 Mains fuse type and ratings 34 Table D 3 XenoWorks BRM Controller cables and recep...

Страница 12: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 4 10 20141208 x This page intentionally left blank...

Страница 13: ...echnical support is provided by Sutter Instrument Company at no charge to our customers Our technical support staff is available between the hours of 8 00 AM and 5 00 PM Pacific Time at 415 883 0128 Y...

Страница 14: ...e working surface and the default axis polarity settings used see Axis Polarity a displacement of the joystick to the left for example will result in concomitant movement of the micromanipulator to th...

Страница 15: ...cromanipulator in the X and Y axes are achieved by moving the joystick left and right and forward and backward respectively the switches for establishing axis polarity for right and left handed manipu...

Страница 16: ...ip the X and Y movement will remain deactivated for as long as your finger is touching the band The link is reestablished by releasing the touchpad See Operating Instructions for more details on how t...

Страница 17: ...manipulator should not be used on the left side of the microscope and vice versa The Home position is a factory preset and cannot be re configured See Operating Instructions for details 1 4 2 9 Speed...

Страница 18: ...this and thus protect the user from inadvertently running out of motor travel during an experiment By following the instructions included in this manual use of the Setup function will position each o...

Страница 19: ...e is inverted The axis polarity controls are located on the rear of the joystick column Each DIP switch control affects one axis switch 1 reverses the X axis switch 2 reverses the Y axis and switch 3...

Страница 20: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 8 This page intentionally left blank 4 10 20141208...

Страница 21: ...l is positioned correctly with respect to the microscope s optical axis Each model microscope will require special Y axis slide settings To determine the appropriate setting for your microscope refer...

Страница 22: ...Slide the dovetail on the bottom of the adapter plate into the slide of the microscope adapter it may be necessary to loosen the locknut on the rear of the microscope adapter Do not tighten the locknu...

Страница 23: ...e end of travel is reached in the X axis Use the Declutch if necessary Take care not to let the mechanical touch any part of the microscope particularly the condenser Ensuring that the mechanical base...

Страница 24: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 12 See the next section for details on how to align a micropipette at the start of an injecting session 4 10 20141208...

Страница 25: ...ith Respect to the Joystick The Declutch mechanism is used to re center the joystick with respect to the micropipette tip This is useful when the micropipette is at the very edge of the microscope fie...

Страница 26: ...tforms The following paragraphs describe factors that affect the setting of the micropipette holder 3 1 3 1 Microinjection Chamber Some types of microinjection chamber plastic culture dishes for examp...

Страница 27: ...ngled micropipettes ICSI Stem Cell Nuclear Transfer a b Figure 3 4 a and b Suggested approach for suspended cells with straight micropipettes Pronuclear Injection a b Figure 3 5 a and b Suggested appr...

Страница 28: ...r either in front or behind the micropipette holder clamp 3 1 4 Memorizing the Work Position This is accomplished by first using the joystick to position the micropipette in the desired Work position...

Страница 29: ...pe 2 Slowly rotate the Z axis control clockwise lowering the micropipette towards the bottom of the injection chamber following the micropipette tip down with the microscope s fine focus Use slower sp...

Страница 30: ...ette is being replaced by one that is longer the micropipette may crash into the bottom of the injection chamber upon returning to Work from Home Several measures taken alone or together can prevent t...

Страница 31: ...the Z axis by rotating the Z axis dial at the bottom of the joystick 8 Once the desired magnification has been reached and the micropipette is still in view set the Work position 9 To replace the micr...

Страница 32: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 20 4 10 20141208 This page intentionally blank...

Страница 33: ...or 1 Change the range of movement setting Speed to 2 3 4 or 5 2 The micropipette is too long Use a micropipette that is no longer than about 50mm There is excessive vibration at the micropipette tip...

Страница 34: ...e of the chamber consider repositioning the micropipette with respect to the micromanipulator as described in Section 4 1 The micropipette holder clamp has not been tightened down sufficiently Tighten...

Страница 35: ...cts the center of the micromanipulator s Y axis Realign the micropipette holder clamp arrangement X and Z axes so that when the micropipette holder is replaced in the clamp the micropipette tip will c...

Страница 36: ...is is usually only noticeable at higher magnifications Angled micropipettes should be adjusted from the start so that they are slightly toe down in other words the tip should touch the bottom of the m...

Страница 37: ...Ensure that the cable connecting the mechanical to the controller is not under tension coil the cable a few times on the bench surface behind the microscope 4 3 Transporting the Micromanipulator When...

Страница 38: ...position 9 Locate the original shipping screws that should have been retained with the original packaging and insert them into the appropriate locations Do not over tighten the screws turn them until...

Страница 39: ...s paid for by Sutter Instrument during the first three months this warranty is in effect after which the cost is the responsibility of the customer The limited warranty is as stated above and no impli...

Страница 40: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 28 4 10 20141208 This page intentionally blank...

Страница 41: ...cal extension X285305 Z axis horizontal extension X285310 Rotating Base Rotating base 225RBI Mounts Hinged headstage mount 285HEA Adapters Mounting adapter plate X285210 For use with MT and MD series...

Страница 42: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 30 4 10 20141208 This page intentionally blank...

Страница 43: ...also stored within the fuse holder The spare fuse is concealed in a compartment as shown To remove the spare fuse press down on the end of the compartment to push it out of the other end The old fuse...

Страница 44: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 32 4 10 20141208 This page intentionally blank...

Страница 45: ...ersion 6 position rotary range of movement control Setup function Position memory One user defined set point Work One user defined axis limit Z floor Two preset points Home and Setup Ranges of Movemen...

Страница 46: ...connectors Controller Rear Panel Port Connector Receptacle Cable Connector Types Connects to Cable Type Cable Max Length Power entry 3 pin male connector 3 pin power standard female 3 pin male Geograp...

Страница 47: ...rly to 67 35 C 95 F Regulatory Safety EN61010 1 CE EMC EN61326 CE Cleaning 70 alcohol or e g UV Dimensions H x W x D Mechanical 112 x 145 x 185 mm 4 4 x 5 7 x 7 3 in Joystick 265 x 240 x 226 mm 10 4 x...

Страница 48: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 3 4 10 20141208 6 This page intentionally blank...

Страница 49: ...d Features Micropipette Holder Clamp 5 Frequently used Functions and Features Swing Gate 6 fuse holder 31 location 31 mains 34 replacement 31 spare 31 fuses replacement mains iii 31 G General Informat...

Страница 50: ...opipette type 16 Piezo impact drive 16 Working distance 16 Setting the Angle of Attack of the Micropipette 14 specifications cleaning 35 dimensions 35 operating environment 35 regulatory 35 weight 35...

Страница 51: ...4 10 20141208 39 NOTES XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV...

Страница 52: ...XENOWORKS MICROMANIPULATOR SYSTEM TOUCHDECLUTCH OPERATION MANUAL REV 40 4 10 20141208 NOTES...

Отзывы: