![Sutter Instrument TRIO MPC-165 Series Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/sutter-instrument/trio-mpc-165-series/trio-mpc-165-series_operation-manual_1407941035.webp)
TRIO MPC-165 SERIES MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.67K2 (20201130) (FW V2.62)
33
Table 4-11. Move to specified “Work” position (‘W’) command.
Tx/-
Delay/-
Rx
Ver. Total
Bytes
Byte
Offset
(Len.)
Value
Alt-key-
pad #
Ctrl-
char
ASCII
def./-
char.
Description
Dec. Hex.
Binary
Tx
All
13
0
87 57 0101 0111 0087
‘W’
Command
1 (4)
X µsteps
5 (4)
Y µsteps
9 (4)
Z µsteps
Rx
All
1
0
13 0D 0000 1101
^M <CR>
Completion indicator
NOTE: All positions are in microsteps (µsteps): 32-bit (4 bytes) positive (unsigned) integer
values, in Little Endian bit order (see Notes).
4.9.8
Move in Straight Line to Specified Position at Specified Speed (‘S’) Command
This command instructs the controller to move all three axes simultaneously in a straight
line to specified position (see
Ranges
table), at a specified speed (1 of 16 speeds). The
command sequence consists of fourteen bytes.
Table 4-12. Straight-line move to specified position (‘S’) command.
Tx/-
Delay/-
Rx
Ver. Total
Bytes
Byte
Offset
(Len.)
Value
Alt-key-
pad #
Ctrl-
char
ASCII
def./-
char.
Description
Dec. Hex.
Binary
Tx
All
14
0
83 53 0101 0011 0083
‘S’
Command
1 (1) 15
-
0
0F
-
00
0000 1111
-
0000 0000
0015
-
0000
^O
-
^@
Speed (15 – 0 (fastest
through slowest))
2 (4)
X µsteps
6 (4)
Y µsteps
10(4)
Z µsteps
Rx
All
1
0
13 0D 0000 1101
^M <CR>
Completion indicator
NOTE: All positions are in microsteps (µsteps): 32-bit (4 bytes) positive (unsigned) integer
values, in Little Endian bit order (see Notes).
While all move commands cause movement to occur at a rate of 5,000 microns/second, the
“Straight-Line Move ‘S’ command is specified with one of sixteen speeds. Actual speed for the
can be determined with the following formula: (5000 / 16) * (sp +1), where 5,000 is the
maximum speed in microns/second and “sp” is the speed level 0 (slowest) through 15
(fastest). For mm/second or microns/millisecond, multiply result by 0.001.