![Sutter Instrument TRIO MPC-165 Series Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/sutter-instrument/trio-mpc-165-series/trio-mpc-165-series_operation-manual_1407941034.webp)
TRIO MPC-165 SERIES MICROMANIPULATOR SYSTEM OPERATION MANUAL – REV. 2.67K2 (20201130) (FW V2.62)
32
4.9.4
Move to Controller-Defined HOME Position (‘h’) Command
This command moves to the position saved by the controller’s HOME button. X & Z move
first (angle determines order and simultaneity), and Y last.
Table 4-8. Move to controller-defined HOME position (‘h’) command.
Tx/-
Delay/-
Rx
Ver. Total
Bytes
Byte
Offset
(Len.)
Value
Alt-
key-
pad #
Ctrl-
char
ASCII
def./-
char.
Description
Dec. Hex.
Binary
Tx
All
1
0 104 68 0110 1000 0104
‘h’
Command
Rx
All
1
0
13 0D 0000 1101
<CR>
Completion indicator
4.9.5
Move to Controller-Defined WORK Position (‘w’) Command
This command moves to the position saved by the controller’s WORK button. Y moves first,
and X & Z last (angle determines order/simultaneity).
Table 4-9. Move to controller -defined WORK position (‘w’) command.
Tx/-
Delay/-
Rx
Ver. Total
Bytes
Byte
Offset
(Len.)
Value
Alt-key-
pad #
Ctrl-
char
ASCII
def./-
char.
Description
Dec. Hex.
Binary
Tx
1
0 119 77 0111 0111 0119
‘w’
Command
Rx
1
0
13 0D 0000 1101
<CR>
Completion indicator
4.9.6
Move to Specified “Home” Position (‘H’) Command
This command instructs the controller to move all 3 axes to specified position, moving X & Z
(angle determines order/simultaneity), and Y last (see
Ranges
table).
Table 4-10. Move to specified “Home” position (‘H’) command.
Tx/-
Delay/-
Rx
Ver. Total
Bytes
Byte
Offset
(Len.)
Value
Alt-key-
pad #
Ctrl-
char
ASCII
def./-
char.
Description
Dec. Hex.
Binary
Tx
All
13
0
72 48 0100 1000 0072
‘H’
Command
1 (4)
X µsteps
5 (4)
Y µsteps
9 (4)
Z µsteps
Rx
All
1
0
13 0D 0000 1101
^M <CR>
Completion indicator
NOTE: All positions are in microsteps (µsteps): 32-bit (4 bytes) positive (unsigned) integer
values, in Little Endian bit order (see Notes).
4.9.7
Move to Specified “Work” Position (‘W’) Command
This command instructs the controller to move all 3 axes to specified position, moving Y first,
and X & Z last (angle determines order/simultaneity) (see
Ranges
table).