Table 2-1. DIP Switch configuration settings.
Switch Definition Position
Definition
ON*
Same as TTL IN:
TTL IN=low
Æ
TTL OUT=low;
TTL IN=high
Æ
TTL OUT=high
1
TTL OUT logic
OFF
Inverse of TTL IN:
TTL IN=low
Æ
TTL OUT=high;
TTL IN=high
Æ
TTL OUT=low
ON
TTL IN=high turns lamp on
2
TTL IN logic
OFF*
TTL IN=low turns lamp on
ON*
Lamp turns on (Notes 1 and 2)
3
Lamp state when switching
from manual control to TTL
OFF
Lamp turns off (Note 1)
ON Lambda
HPX-L5
4
Lambda HPX model-specific
operation (Note 3)
OFF*
Lambda HPX
“*” = Default
NOTE 1: After switching from manual mode (LED switch position ON or OFF) to TTL mode
(LED switch position TTL), TTL control of the lamp’s on/off state will resume with the next
change of logic state at TTL IN.
NOTE 2: While using the LED switch to switch from ON to TTL, the lamp turns off while
transitioning through the OFF position (the OFF position is located in between ON and
TTL).
NOTE 3: Switch 4 must be OFF for proper operation of the Lambda HPX (liquid-cooled