LAMBDA 721 OPERATION MANUAL – REV. 1.10C (20210517)
31
APPENDIX C.
FUSE REPLACEMENT
In the event that the lamp fails to power up when the power switch is turned on, please
check to see if the line power fuse has blown. The fuse is located in a pry-out holder on the
power entry module on the back of the unit. To remove the fuse holder first unplug the
power cord from the power entry module. This will reveal a slot just under the edge of the
fuse holder. Use a screwdriver to pry the holder straight out of the power entry module.
Figure C-1. Power entry module and fuse detail.
The fuse holder holds two fuses. Both fuses are of the same type and rating. If either fuse is
blown, it is recommended that both fuses be replaced.
The type and rating of both fuses are as follows:
5 x 20 mm glass tube, Time Lag (IEC 60127-2, Sheet 6)
T3.15A 250V (Time Lag, 3.15 Amps, 250 Volts)
(Littelfuse 02193.15MXAP (219XA Series, RoHS compliant))