30
Command
Description
Command Syntax
(command underlined)
ACCUBAR
®
response (underlined)
"a" represents the
single-character address
M6
Request
Temperature
and Pressure
Measurement
(Version 2.0
and higher)
aM6!
aD0!
attt4
ttt is the time in seconds until the
measurement is ready and 4 is the
number of values that can be collected
a
service request
atupv
where t is the temperature, u is the
temperature units, p is the pressure, and
v is the pressure units. Use the XUT
command to set the temperature units
and the XUP command to set the
pressure units.
M7
Request
factory
calibration
Pressure and
Temperature
Measurement
(Version 2.0
and higher)
aM7!
aD0!
attt2
ttt is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
a
service request
apt
where p is the pressure in mB and t is the
temperature in degrees Celsius.
MC
Request
Default
Pressure
Measurement
with CRC-16
(Version 2.0
and higher)
aMC!
aD0!
attt2
ttt is the time in seconds until the
measurement is ready, 2 is the number of
values that can be collected
a
service request
axuC
where x is the signed pressure value, u is
the signed indicator of the units, and C is
the 3 character CRC. The units are set
by the XUP command.
MC1 Request
Pressure
Measurement
in mB with
CRC-16
(Version 2.0
and higher)
aMC1!
aD0!
attt1
ttt is the time in seconds until the
measurement is ready, 1 is the number of
values that can be collected.
a
service request
apC
where p is the signed pressure value in
mB and C is the 3 character CRC