Bringing the Benefits of Real-Time Data Collection to the World
Sutron Corporation, 22400 Davis Drive, Sterling, Virginia 20164-4444
25
Command
Description
Command Syntax
(command underlined)
Sensor response (underlined)
"a" represents the
single-character address
C2
Request
Concurrent
Temperature
Measurement
(Version 2.0
and higher)
aC2!
aD0!
attt02
ttt is the time in seconds until the
measurement is ready and 02 is the
number of values that can be collected
atu
where t is the temperature and u is the
units 0= Celsius and 1=Fahrenheit. Use
the XUT command to set the units.
C3
Request User
Scale, User
Offset, and
Field
Calibration
Offset
(Version 2.0
and higher)
aC3!
aD0!
a00003
000 is the time in seconds until the
measurement is ready and 03 is the
number of values that can be collected
asoc
where, s is the user scale and o is the
user offset (mB), and c is the field
calibration offset (mB).
C4
Request
Standards lab
Calibration
Scale and
Offset
(Version 2.0
and higher)
aC4!
aD0!
a00002
000 is the time in seconds until the
measurement is ready and 02 is the
number of values that can be collected
aso
where, s is the scale calibration and o is
the offset calibration(mB).