36
Care and Maintenance
EN
9
Care and Maintenance
Special care is necessary when servicing, maintaining, and storing the device and
its accessories to maintain the functionality of the device and its accessories.
Authorized trained personnel
Only a SurgiQuest technician may perform repairs, adjustments, or alterations
on the device or accessories and use the service menu. Any violation will void the
manufacturer's warranty.
9.1
Cleaning the Device
Manufacturer’s specifications
The manufacturer recommends that the device is cleaned after every procedure
as follows:
1. Use the
ON/OFF
key to turn the device off.
2. Remove the power cable.
3. Wipe all outer surfaces of the device with a disposable soft cloth moistened
with an appropiate disinfectant (for example Meliseptol® rapid). The concen-
tration of the disinfectant and the application should be adjusted according
to the information provided by the manufacturer of the disinfectant. Make
sure moisture does not enter the device.
NOTE!
Do not sterilize the device.
9.2
Maintenance Intervals
Manufacturer's specifications
The manufacturer stipulates that qualified personnel or hospital technicians
must regularly test the device to assess its functionality and technical safety. The
device needs to be serviced at least every two years. The tests are described in
chapter 10 "Annual Inspection".
Regular inspections will assist in early detection of possible malfunctions. This
helps preserve the device and increases its safety and service life.
NOTE!
Service or maintenance work may not be carried out during surgery.
9.3
Maintenance by Authorized Service Technician
Two-year maintenance interval
An authorized service technician has to inspect and service the device at appro-
priate intervals to ensure the safety and functionality of the unit. The minimum
service interval is two years, depending on frequency and duration of use. If the
service interval is not maintained, the manufacturer does not assume any liabil-
ity for the functional safety of the device.
A sticker located on the rear panel of the device will remind you of the latest date
for the next service or maintenance check.
Authorized service technicians are only trained and certified by the manufactur-
er.
Authorized trained personnel
All of the service tasks, such as changes, modifications, repairs, calibrations, etc.
may be carried out only by the manufacturer or manufacturer-approved trained
and skilled technicians.
Unauthorized personnel
The manufacturer is not liable for the operational safety of the device if unau-
thorized persons conduct this maintenance or any other service tasks.
Liability
Unauthorized opening of the device and repairs performed by unauthorized per-
sonnel or third parties and/or changes or modifications release the manufactur-
er of any liability concerning the operational safety of the device.
Technical documents
Receiving technical documentation from the manufacturer does not authorize
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...