Annual Inspection
39
EN
10
Annual Inspection
Measured values and tolerances
Each test conducted should be documented with date and signature on the test
log.
The following measuring tools and resources were used by the manufacturer to
determine the listed measurements and tolerances:
An authorized service technician must check the device if the specified parame-
ters and tolerances are exceeded.
10.1
Electrical Safety Test
1. Perform a visual inspection. Make sure that
• the fuse corresponds with the specifications indicated by the manufactur-
er,
• labels and stickers on device are legible,
• the mechanical condition of the device allows for its safe use,
• the device is clean to ensure proper and safe functionality.
2. Carry out the measurements for the ground leakage current, short-circuit
current/housing leakage current, and the protective conductor resistance as
per IEC 62353 in the current version or according to the applicable national
standard.
10.2
Basic Function Test
1. Conduct this test without the Single Lumen Filtered Tube Set connected to
the AirSeal® iFS
2. Use the
ON/OFF
switch to turn the device on. The device now conducts a self-
test. A short acoustic signal can be heard. Press the
Standard Insufflation
Mode
key to select this operating mode.
3. The factory default settings are 15 mmHg for the nominal pressure and 5 l/
min for the nominal flow.
4. The following values are displayed:
Nominal pressure 15 mmHg
Nominal gas flow rate 5 l/min [button 1]
Actual pressure 0 mmHg
Gas consumption 0 l
5. Start insufflation:
Press the
START
key.
The following values are displayed: Actual pressure 0 mmHg
Initial insufflation
is depicted.
Streaming gas can be heard at the insufflation tube connection.
6. Select the max. nominal gas flow.
The following values are displayed:
Actual gas flow equals nominal gas flow rate
Actual pressure 0 mmHg
Initial Insufflation
is displayed.
Streaming gas can be heard at the insufflation tube connection (barbed con-
Manometer
Range 0-100 mmHg, error class ± 1.6%
Syringe
60 ml
Silicone tube
8 mm x 2 m
T adapter
8-8-8 mm
Veress needle
length 100 mm
opening diameter 1.4 mm,
inner cannula diameter 1.6 mm
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...