6.4.3 Load Calibration (LoAd)
After doing Data Calibration, if there are conditions for Load Calibration, do Load Calibration with
loading standard weight on the weigher to correct Span Coefficient further for higher weighing accuracy.
The loading weight should be bigger than 50% of Scale Capacity value.
Input ‘Calibrating Weight’: 1~999999 Weight Unit.
Main Display Interface
【
■
】
+
【
►
】
【
▲
】
: F2-CAL
【
】
+
【
►
】
【
▲
】
: -LoAd-
【
】
M
【
■
】
: Exit
【
】
: Save
M
【
】
【
►
】
【
▲
】
:
105: New Span Coefficient value.
oLd: Original Span Coefficient value.
Sr: Span Correction Ratio = New Value / Original Value.
Ad: AD Value.
If AD Value
≤
Zero Value, display ‘Err’, and it’s not allowed to save
Span Coefficient.