14
6.2 Main Menu
Main Menu
Second Menu
Sign
Function
Sign
Description
F1.SEt
Parameter
Setting
-SCAL
Scale parameters setting.
-CALP-
Calibration parameters setting.
-SEtP-
Setpoint parameters setting.
-SErP-
Communication parameters setting.
-dISP-
User interface parameters setting.
F2.CAL
System
Calibration
-ZEro-
Zero Calibration without loading on the weigher to correct Zero
Value.
-dAtA-
Data Calibration: Input the specification parameter values of
loadcell according to the actual configuration of the weighing
system to correct Span Coefficient. If there is no access to get the
specification parameter values for Data Calibration, then it’s
necessary to do Load Calibration.
-LoAd-
Load Calibration: After doing Data Calibration, if there are
conditions for Load Calibration, do Load Calibration with loading
standard weight on the weigher to correct Span Coefficient further
for higher weighing accuracy.
-SEgC-
Segmenting Span Correction: After doing Zero Calibration and
Load Calibration [or Data Calibration], it’s allowed to do
Segmenting Span Correction with loading standard weight on the
weigher by 10 Correction Points to get Span Correction Ratio of 10
linearity segments for higher weighing accuracy.
F3.rEC
Unused.
F4.CLn
Clear Screen: Clear the display value of Peak Hold Value and
Weight-checking Value.
F5.Loc
Key-locker
-oPEn-
Key-unlocking.
-Locc-
Key-locking.
-PASS-
Password Set.
Exfactory Passwords:
Operator Password: 000.
Administrator Password: 001.