Chapter 5 Configuration
www.supmea.com
14
Pressure setting: setting pressure under current altitude, default
atmosphere pressure 760mmHg
。
Salt Content Setting: set the salinity of the current solution. The default is
0.00ppt
Temperature unit: available to choose which temperature unit to display, by
default is
℃
.
DM2800
-------------------- Signal Setting --------------------
1. Unit switch
2. Salt Content Setting
3. Air Pressure Setting
4. Temp Switch
5. Electrode Switch
Unit switch: Switch among mg/L, % and hPa; mg/L for dissolved oxygen
value; % for saturation value ; hPa for oxygen partial pressure.
Salt Content Setting: Set the salinity of the current solution, range 0.00
~
99.99g/Kg
。
Air Pressure Setting: set the air pressure value of the current altitude. The
setting range is 500 - 1100 hPa.
Temp Switch: Set the automatic temperature compensation NTC 10K or
PT1000 and manual temperature compensation value, and the measuring
range is -10 - 60
℃
.
Electrode Switch: It can switch between DS120 and DS130 coated
dissolved oxygen sensors.