Chapter 2 Installation
www.supmea.com
7
DM2800
Identification of terminal
DO+: Dissolved oxygen electrode anode
DGND: Dissolved oxygen electrode shield wire
DO-: Dissolved oxygen electrode cathode
A: Temperature compensation terminal A, NTC10K A or PT1000 A
B: temperature compensation terminal B, NTC10K B or PT1000 B
C: Temperature compensation terminal C, short-circuit terminal, PT1000
three-wire system, short-circuit of PT1000 two-wire system to B, NTC10K
does not need to be connected to C
485A+: RS485 communication output terminal A+
485B-: RS485 communication output terminal B-
I+: (4
~
20)mA output t
I-: (4
~
20)mA output terminal-
HO: High alarm normally open relay
HC: High alarm normally closed relay
COM: Common terminal
LO: Low alarm normally open relay
LC: Low alarm normally closed relay
COM: Common terminal
220V N: AC 220V neutral line
NC: Null