Chapter VI Calibration of MLSS/TSS
20
Chapter VI Calibration of MLSS/TSS
The MLSS/TSS sensor has been factory calibrated before delivered, and
the self-calibration can be carried out according to the following steps if
required. The use of standard liquid is required for the calibration of the
MLSS/TSS , taking two-point calibration as an example:
1. Connect the sensor to the transmitter.
2. Set up the related parameters (set the MLSS/TSS factoras1 in the
signal setting), enter the MLSS/TSS Calibration actual value measurement
interface and clean the sensor.
3. Put the probe into the first point standard liquid (usually take the pure
water as the first point). After the data is stable, read the actual value of the
measurement and record the data.
4. Remove the probe from the first standard liquid, rinse it with clean
water, and wipe the sensor.
5. Put the probe into the second standard liquid. After the data is stable,
read the actual value of the measurement and record the data.
6. Click OK after recording the data; enter the standard selection
interface and select the two-point calibration; enter the first target value
setting interface.
7. Enter the target value of the first standard liquid measured at the
laboratory and click OK; enter the first point actual value setting interface and
then enter the first point actual value recorded in step 3; click OK to enter the
second point target value interface.
8. Enter the target value of the second standard liquid measured at the
laboratory and click OK; enter the second point actual value setting interface
and then enter the second point actual value recorded in step 3; click OK to
return the main menu.