Chapter IV System menu
15
Alarm Setting
Figure 15
TUR High Alarm: When the measured value is larger than the high
alarm pick-up value, the high alarm relay is picked up and when the measured
value is less than the high alarm breaking value, the high alarm relay is
disconnected.
TUR Low Alarm: When the measured value is less than the low alarm
pick-up value, the low alarm relay is picked up and when the measured value
is larger than the low alarm breaking value, the low alarm relay is
disconnected.
MLSS High Alarm: When the measured value is larger than the high
alarm pick-up value, the high alarm relay is picked up and when the measured
value is less than the high alarm breaking value, the high alarm relay is
disconnected.
MLSS Low Alarm: When the measured value is less than the low alarm
pick-up value, the low alarm relay is picked up and when the measured value
is larger than the low alarm breaking value, the low alarm relay is
disconnected.