Chapter 2 Starting Measurement
www.supmea.com
13
*standard pipe materials and standard liquids refer to those with the sound
parameters that have already been programmed into software of the flow meter,
therefore there is no need to configure them
(4) Liner material and its sound speed and thickness, if there is any liner.
(5) Liquid type (for non-standard liquids, the sound speed of the liquid is also
needed)
(6) Transducer type adapted to the flow meter. Generally the standard Ml or the
Frame M-sized transducers will be the selected option.
(7) Transducer mounting methods (the V-method or Z-method is the common
option)
(8) Check up the space displayed on M25 and install the transducers accordingly.
For standard pipe materials and standard liquids, the following detailed
step-by-step set up is recommended.
(1) Press keys
to enter M11 window to input the digits for the pipe
outer diameter, and then press
key.
(2) Press key
to enter M12 window to input the digits for the pipe outer
diameter and then press
key.
(3) Press key
to enter M14 window, and press
key to enter the option
selection mode. Use keys
and
to scroll up and down to the intended
pipe material, and then press
key.
(4) Press key
to enter M16 window, press
key to enter the option selection
mode, use keys
and
to scroll up and down to the liner material, and then
press
key. select "No Liner", if there is no liner.
(5) Press key
to enter M20 window, press
key to enter the option selection
mode, use keys
and
to scroll up and down to the proper liquid, and then
press
key.
(6) Press key
to enter M23 window, press
key to enter the option
selection mode, use keys
and
to scroll up and down to the proper
transducer type, and then press
key.
(7)Press key
to enter M24 window, press
key to enter the option
selection mode, use key
and
to scroll up and down to the proper transducer
mounting method, and then press
key.