53
System Errors
ERR Message
If an error message appears on the display in the format “
ERR=XX XX XX XX
,” the
radon monitor experienced an unusual system error. Normally this indicates a
hardware or firmware failure. Make a note of the 8-digit error code and contact a
Sun Nuclear Support representative for assistance.
Notice Message
If the message “NOTICE: Press 5” appears in the display upon waking up, there
has been an error.
1
To read the error, press
5
. The display:
•
Briefly displays “ERROR Stopping Test”
•
Then displays “Read using PC”
•
Then displays “Start Test = 5”
However, the error condition has not been cleared.
2
Connect device to computer and launch Radon Monitor application.
3
On the menu, click
Com Port
and verify the correct Com port is selected.
4
Select
View > View Notices
, then
Read Notices from Monitor
. Make note
of the error explanation that displays on screen. This information may be
useful to Sun Nuclear Support.
5
Select
Save to Log
and save the error code text to a log file.
6
Select
Clear Notices from Monitor
to remove the error code from the
device.
7
Download measurement data and continue with test.
8
If error condition continues, an error message appears again after the test
and must be cleared again. If this occurs, device likely requires service. See
Contacting Sun Nuclear Support
Service and Calibration
For service or calibration, the unit must be returned to Sun Nuclear Corporation.
The recommended calibration frequency for the Model 1028 or 1029 radon
monitor is one year. See your State or Proficiency listings for requirements.
Note:
Firmware version 0102 can cause a data error of 00 00 00 40
when attempting to print a report with a graph when no data is in the
radon monitor. If this occurs, please call Sun Nuclear Support. See
Contacting Sun Nuclear Support on page 54.
WARNING:
The unit contains high-voltage circuits. Do not open the
case. There are no user-serviceable parts inside.