15
Clearing a Reset
1
Turn on the radon monitor to check the battery. If the system is accidentally
reset, clear the reset as shown in the table below.
2
Start the radon monitor again. See
3
In Setup, change the date and time. See
Low Battery Messages
A 9V battery must always be installed, even when using an AC power supply, to
prevent discharge of the internal 3V battery. When the battery is low or needs to
be replaced, a message is displayed on the radon monitor. The following sections
describe the messages that may be displayed.
Low Battery with AC Power Connected
If the battery is low and the AC power adapter is connected, a message appears
when key 5 is pressed.
Replace the 9V battery as soon as the test is complete. See
Press…
Display Shows…
Comment
Battery OK
Battery is checked.
–
System has reset
System has been reset.
–
Check Date/Time
An alert to check the date and time and re-enter
them.
Clear Reset? Y=5
Sets the system back to operational mode.
–
Shutting down...
System shuts down.
Note:
A reset does not clear or change any test data stored in memory.
Press…
Display Shows…
Comment
Low/No Battery!
Battery is checked. Does not have enough power
for a 60-hour test.
–
Replace Battery!
Replace the 9V alkaline battery as soon as conve-
nient. Current test can continue until complete if AC
power is reliable.
–
Shutting down...
The system shuts down to save power.