DURING TREATMENT

If these suggestions do not resolve the ear pressure, open the red relief 
knob until the feeling of pressure is gone. Then close the red knob to 
continue to pressurize the chamber. The red relief knob will allow you to 
control the speed of pressurization to aid in equalizing the pressure on 
your ears.

Содержание SHALLOW DIVE

Страница 1: ...SHALLOW DIVE...

Страница 2: ...on choosing one of the most innovative and safe hyperbaric chambers In addition to this manual we have an easy to follow instructional video available https www summit to sea com instructions Click o...

Страница 3: ...other on the market Now that you have all the equipment to begin your treatments it is extremely important to watch this entire video and read the manual that was included on a thumb drive in your pac...

Страница 4: ...est all of our chambers to ensure that Summit to Sea chambers have the highest quality comfort and safety We have implemented 4 safety features that are unique to Summit to Sea chambers Dual compresso...

Страница 5: ...pressors fails you would still receive ample fresh air delivered into the chamber allowing you to deflate the chamber and safely exit Summit to Sea uses these two independent compressor systems which...

Страница 6: ...ves When you are at full pressure air will release from the blue valves Your gauge needle should be in the blue range The yellow valve is a safety back up valve set at a slightly higher pressure which...

Страница 7: ...T Zipper stress relief system buckles It is very important that when operating the chamber all belts are buckled and both zippers are completely closed The belts reduce stress on the zippers extending...

Страница 8: ...hyperbaric industry whose chamber system will alert you to a power outage In the event that there is a power outage which would prevent the compressors from supplying fresh air into the chamber the us...

Страница 9: ...hambers are Class II medical devices cleared by the FDA requiring a prescription As we are constantly developing new technology to improve our chambers there may be a photo or video footage that doesn...

Страница 10: ...r with carry bag and two cradles Frame 2 Compressors In the OPEN ME FIRST BOX Silicone and Zipper cleaner Auxiliary Valve Thumb drive with set up instructions Failed Circuit Alarm with battery include...

Страница 11: ...ONLINE MANUAL If you ever misplace this thumb drive you can log on to www summit to sea com instructions to find you chamber s manual...

Страница 12: ...mpressors to the electric outlets and the chamber 2 Use of Summit to Sea s unique zipper system 3 Inflating the chamber 4 Connecting the stabilizing cradles to the chamber 5 Connecting the chamber to...

Страница 13: ...compressors to the electrical outlet Please note that some wall outlets are configured so that one of the sockets is connected to a light switch Avoid using any of these outlets as they could result o...

Страница 14: ...compressors to the wall outlets Using a power strip Using a remote switch Using both the power strip and remote switch The following diagrams will describe several options for connecting your compress...

Страница 15: ...OMPRESSOR 2 SET UP COMPRESSORS CONNECTED TO THE WALL OUTLET Plug the circuit alarm into the wall outlet Next plug one of the compressors into the back of the circuit alarm plug Then plug the second co...

Страница 16: ...OMPRESSOR I I I I I I I I I I POWER STRIP SET UP USING A POWER STRIP Plug the circuit alarm into the wall outlet Next plug the power strip into the back of the circuit alarm plug Now you can plug your...

Страница 17: ...2 REMOTE CIRCUIT ALARM COMPRESSOR 1 SET UP USING A REMOTE SWITCH Plug the circuit alarm into the wall outlet Next plug the remote into the back of the circuit alarm plug Now you can plug your compress...

Страница 18: ...REMOTE I I I I I I SET UP USING BOTH THE POWER STRIP AND REMOTE SWITCH Plug the circuit alarm into the wall outlet Next plug the remote into the back of the circuit alarm plug Then plug the power stri...

Страница 19: ...Remove the chamber from the carry bag Lay out the chamber so the zipper is facing up Summit to Sea provides a unique two zipper combination one internal air tight zipper and one external mechanical z...

Страница 20: ...the two sides of the buckle together until you hear the click It is important to connect the buckles before closing the zippers This allows the zipper to line up making it easier to zip and taking str...

Страница 21: ...a few inches realign then slowly continue to close it Notice the zipper has a dock at each end of the zipper You can apply zipper lubricant on the dock to ease the closure When the internal zipper pul...

Страница 22: ...SET UP The picture on the left shows the slider completely closed into the dock If it is not completely closed the chamber will not fully pressurize RIGHT WRONG...

Страница 23: ...SET UP Next close the external zipper To be sure it is fully closed pull the zipper string through the safety strap You will not see the slider when it is fully closed...

Страница 24: ...SET UP Now that you have both zippers completely closed make sure the red deflation knob is in the closed position...

Страница 25: ...onnectors are located on the back of the chamber Simply click the hose onto the fill valves It is recommended to put a small dot of silicone lubricant on the O ring on the end of each fill valve conne...

Страница 26: ...is to shape connect the stabilizer cradles to the bottom of the chamber Lift the chamber one end at a time sliding the stabilizer cradles under the chamber and connect them to the velcro patches that...

Страница 27: ...SET UP Now that the stabilizers are attached you will build the frame around the chamber You will notice the chamber has velcro loops at each end to attach the white frame...

Страница 28: ...SET UP For you to better understand how the frame will look when it is built this picture shows the frame without the chamber...

Страница 29: ...SET UP End pieces Floor poles Top pole with end caps Arches...

Страница 30: ...SET UP Place the end pieces at each end of the chamber Next place the arches into the end pieces...

Страница 31: ...SET UP Place the end pieces at each end of the chamber Next place the arches into the end pieces Make sure the screws are facing towards the center...

Страница 32: ...SET UP Two of the poles have end caps These will screw together and go over the top of the chamber...

Страница 33: ...UP Screw the other 2 pairs of poles without end caps together Place one of the poles in front of the chamber and insert it into the connector on the end pieces Repeat this on the back side of the cham...

Страница 34: ...UP Screw the other 2 pairs of poles without end caps together Place one of the poles in front of the chamber and insert it into the connector on the end pieces Repeat this on the back side of the cham...

Страница 35: ...SET UP Attach the last pole to the 2 arches using the included screwdriver Do not over tighten the screws...

Страница 36: ...SET UP Now connect each velcro strap to the frame overlapping no more than 1 Do not connect the zipper handle loops to the frame...

Страница 37: ...SET UP Your chamber should look like this Do not put the frame through the zipper loops...

Страница 38: ...et up turn the red deflation knob to release the air and deflate the chamber It is designed to turn in either direction to release air from the chamber As you turn the knob you will notice that you ha...

Страница 39: ...SET UP It is important that you do not open the zippers unless the gauge reads zero...

Страница 40: ...ATING With the buckles still connected unzip the outside zipper Reach under the white strap to grab the pull tab Hold the black zipper loop and pull the zipper open Be sure to open the zipper all the...

Страница 41: ...DEFLATING Next you will open the inside zipper Pull the zipper tab on the inside zipper Open the zipper all the way...

Страница 42: ...DEFLATING Next unbuckle the belts...

Страница 43: ...SET UP Now that you have deflated the chamber you can place the mat inside...

Страница 44: ...closed Do not open the zipper until the gauge reads zero Each belt must be fastened from the start of inflation to full deflation The compressors must be running during the whole time of treatment The...

Страница 45: ...ngratulations you have successfully set up your chamber Now we would like to review the operation of the chamber and give you a more detailed explanation of the individual components on your chamber a...

Страница 46: ...n valve is closed The zipper can also be closed from the inside as our zippers have zipper pulls on both sides When it is time to end your treatment you will turn the red deflation valve to deflate th...

Страница 47: ...There are two pressure relief valves with blue circles These are designed and factory calibrated to equalize the pressure at 1 3 ATA or 4 4 PSI They are located at each end of the chamber Pressure Rel...

Страница 48: ...ere is a safety backup pressure relief valve with a yellow circle which is set to open at a slightly higher pressure than the blue valves There may be a slight venting of air which is normal Pressure...

Страница 49: ...ero Once the chamber is fully pressurized air will be released out of the blue pressure relief valves and the needle will appear in the blue area If the needle is anywhere within the blue area you are...

Страница 50: ...COMPONENTS Notice that the gauge is reading zero when you begin It will then reach the blue section on the gauge which is 1 3 ATA You will hear air being released from the blue pressure relief valves...

Страница 51: ...COMPONENTS On the end of the chamber are two fill valves Put a small amount of zipper lubricant on each fill valve O ring...

Страница 52: ...notice that you have control of the speed of deflation Red Deflation Valve This valve can also be used during pressurization to customize the speed of inflation for greater comfort and to aid in the...

Страница 53: ...COMPONENTS Additional Ports There are also two additional ports One at each end of the chamber These are for additional accessories...

Страница 54: ...Valve Next to the window is an auxiliary port where the auxiliary valve attaches This small auxiliary valve is located in your packet in a small red envelope along with the zipper cleaner zipper lubri...

Страница 55: ...ice added pressure on the ears This feeling indicates the need to equalize the pressure The pressure feeling is similar to that of landing in an airplane This pressure is relieved through the Eustachi...

Страница 56: ...maneuver Pinch your nose close your mouth and try gently blowing out your nose Repeat every time you begin to feel that pressure Drinking water Yawning or opening your mouth wide will also help Lastly...

Страница 57: ...ar pressure open the red relief knob until the feeling of pressure is gone Then close the red knob to continue to pressurize the chamber The red relief knob will allow you to control the speed of pres...

Страница 58: ...commended to change the air filters on your compressors every 6 months To order these filters go to www summit to sea com accessories To better understand how to change your filters go to www summit t...

Страница 59: ...ake sure all buckles are closed and both the inside and the outside zippers are completely closed every time the chamber is in use Failure to do this may result in inability to pressurize or damage th...

Страница 60: ...lue relief valves do not release air when it is at full pressure stop your treatment and call your distributor It is important that the relief valves are clear of all obstructions inside and out durin...

Страница 61: ...CAUTION The compressors must remain on for the entire treatment This ensures the circulation and exchange of fresh air to provide a safe and comfortable treatment...

Страница 62: ...CAUTION Do not open the zippers or unbuckle the straps until the gauge reads zero...

Страница 63: ...of the best chambers on the market and you will be totally pleased with the ease of use and the quality of craftsmanship Summit to Sea also believes that we have the best customer service of any hype...

Страница 64: ...contact your distributor first If you have any additional questions please visit our website at www summit to sea com On our home page scroll to the bottom click on the word Support and fill out the...

Страница 65: ...DURING TREATMENT WHAT TO WEAR You may wear any type of clothing in the chamber however comfortable and soft clothing is typically used Do not wear shoes inside the chamber...

Страница 66: ...Feel free to bring food drinks books and personal computers or any other form of entertainment into the chamber It may also be helpful to have your cell phone with you We ask that you refrain from br...

Страница 67: ...ecommended that the compressor filters are changed approximately every 6 months To learn how to change these filters go to www summit to sea com filters To order these replacement filters go to www su...

Страница 68: ...ompressor Filter To see a video on how to change the filters go to www summit to sea com filters There is a filter right under the top plate of the compressor It is recommended that your replace the f...

Страница 69: ...this filter It is recommended that you replace the filter every 6 months Lay the compressor on its side Push the black bulb up and give it a quarter turn to unlatch then pull down Unscrew the silver...

Страница 70: ...see a video of how to reset the safety switch go to www summit to sea com safetyswitch First you will need to remove the compressor cover by loosening the 4 screws with a pair of pilers and then using...

Страница 71: ...CHAMBER CARE Resetting the Safety Switch If the safety switch has been tripped you will see this little red button Move the white switch to the center of the compartment...

Страница 72: ...pper triggers are secured It is necessary to keep this dock lubricated Squeeze a small amount of the silicone onto your finger and spread it out on the dock You will need to do this when it becomes di...

Страница 73: ...IPPERS The outside zipper requires very little maintenance A bottle of zipper cleaner lubricant has been included in your packet to use on a monthly basis Periodically wipe the outside open zipper wit...

Страница 74: ...etroleum based household cleaner An anti bacterial cloth has been included with your chamber To order these cloths go to www summit to sea com accessories FITTINGS HOSES Regularly check the fittings a...

Страница 75: ...ld cleaner Carefully fold the chamber ensuring all gauges and fittings are protected with bubble wrap or a similar material Place the chamber in the carry case and store in a dry area between 40 and 8...

Страница 76: ...d sole discretion will repair or replace the warranty item The warranty does not cover damage as a result of misuse or modification Shipping to and from Summit to Sea is the customer s responsibility...

Страница 77: ...ytime within the two year warranty period The extended warranties also cover all components on the chamber including zipper windows seams valves hoses etc Contact your distributor 1 Year Extended warr...

Страница 78: ...liance i e lamp in the same outlet If the appliance does not work use a different outlet or call an electrician If the appliance does work see a short video on resetting the safety switch To order the...

Страница 79: ...zipper to make sure both ends of the zipper are completely sealed and placed inside the dock If your chamber is not getting to full pressure check to see if the red knob is closed and no air is escap...

Страница 80: ...who may be operating the chamber needs to watch this video By doing this it will ensure a safe environment for all who will be benefiting from the use of the chamber Remember if you have any question...

Страница 81: ...is chamber without guidance and instruction from your health care provider Do NOT use the chamber if you have the following conditions Common colds or flu like symptoms Recent alcohol consumption Bloc...

Страница 82: ...the chamber CAUTION Inspect all seams and fittings prior to use to ensure they are secure Failure to do so could cause leaks reducing the performance of the chamber CAUTION If any unusual noise or od...

Страница 83: ...n valves These are factory set and there are no user adjustments Any sign of tampering with these valves will void the warranty WARNING Do not use the chamber in a manner where the chamber could roll...

Страница 84: ...For proper operation the air compressor is to be running throughout the entire treatment This is important to ensure that CO2 build up does not occur These valves exhaust the old air while maintainin...

Страница 85: ...t With its small carbon footprint this compressor will easily integrate into your environment Oil less 120 with a flow rate of 120 liters per minute See compressor manual included in the compressor bo...

Страница 86: ...ipper An airtight zipper securely seals the air pressure inside the chamber without the need to massage or manipulate the seal along with a heavyduty zipper to ensure the integrity of the chamber With...

Страница 87: ...towel EARS Be sure to chew gum or swallow often during the pressurization ENJOY a beverage read a book or enjoy any other form of entertainment UNZIP the chamber and inspect for foreign objects ENSURE...

Страница 88: ...m 30 60 minutes or longer Check with your medical provider for recommended duration Allow 5 10 minutes for the chamber to return to normal pressure As air is released check with the occupant to ensure...

Страница 89: ...ons The chamber has been cleared by the FDA for the treatment of Acute High Altitude Sickness FDA WARNING Increasing the percentages of oxygen content in the air delivery is considered the same as adm...

Страница 90: ...To order additional filters zipper lubricant zipper cleaner www summit to sea com accessories To learn how to change both filters www summit to sea com filters If you have questions or need technical...

Страница 91: ...SUMMIT TO SEA PO Box 339 Mound MN 55364 Phone 1 877 PSI DIVE 1 877 774 3483 Fax 1 877 420 7622 www summit to sea com 2011 Summit to Sea LLC...