9.Troubleshooting
|
Error message list
9. Troubleshooting
Error message
Possible cause
Troubleshooting
Refer to
There is a foreign object in
the screen.
Clean the microscope lens protection
glass and LEDs.
▸
P.3-2
Cleaning LEDs,
P.3-3 Cleaning
lens protection
glass
▸
Continued from
the previous page.
The splicer failed
to inspect the right
and left fibers.
There is dust on the fiber
end face.
Cleave the left (right) fiber again.
▸
Instruction
manual that
came with
the cleaver
you use.
The splicer is
unable to start
splicing because
the hood is open.
Close the hood.
Unstable arc
(Splice)
The electrodes are installed
incorrectly in the splicer.
Make sure the electrodes are
installed correctly in the splicer.
<Correct>
<Incorrect>
▸
P.3-6, 3-7
Electrode
replacement
procedures