SF-8200 Operation Manual
25
/
76
Click the reagent hole to be unloaded
Click the "Uninstall" button, the hole reagent information disappears, the icon becomes blank;
Remove the reagent bottle corresponding to the hole in the instrument and complete the reagent
unloading manually.
3.2.2. Automatic operation
Automatic loading / unloading of reagent functions is enabled or disabled through the "Devices"
tab of the "Settings" screen. Refer to Chapter 7 "4.3. Reagent Position Detection" for details.
1) Automatically load reagents
2
、根据试剂类型放置在相应区域
1
、扫描试剂条码
In the reagent "reagent" interface, the instrument will automatically open the reagent barcode
scanner;
scan reagent bottle label bar code;
Within 5 seconds, put the reagent into the corresponding area, the system can automatically iden-
tify and set the reagent information;
Click the "confirm" button to complete the reagent loading;
2) Automatic unloading of reagents
remove the appropriate location of the reagent to complete unloading;
Note! The first reagent loading reagent need to enter the reagent information, the specific
operation, please refer to "Chapter 4 1. Batch Management."
4 CHECK THE CALIBRATION CURVE
Make sure the calibration curve is set correctly before proceeding with the project test.
1. Scan the Bar Code of Re-
agent
2. Put the Reagents to corresponding positions
1. Scan the Bar Code of Reagent