SF-8200 Operation Manual
19
/
76
1.2 DRAINAGE INSPECTION
废液满液报警接口
及排水水嘴接头
清洗液缺液报警接口
及进水水嘴接头
Alarm interface with quick connector, insert, you need to limit the connector and the interface of the
instrument alignment;
When the water inlet and exhaust nozzles are equipped with quick-closing faucets, press the reed
of the water faucets, the water faucets will pop up automatically when the faucet is clicked; if it cannot be
inserted, please check the spring Whether the film is pressed in place.
Note! Drainage faucet if not in place, will cause drainage blockage, cannot be discharged in
time, leading to overflow within the instrument.
Check the cleaning liquid tank. If the liquid level of the cleaning liquid is low, replace the liquid
cleaning liquid. The liquid cleaning liquid cannot be replaced by other liquid, and dispose of the waste
liquid tank at the same time.
WARNING: Waste drums contain potentially biodegradable materials. Wear rubber gloves
and dispose of in accordance with local regulations.
Liquid-overflow switch
Liquid deficient switch
Waste liquid barrel
Cleaning liquid barrel
Liquid-overflow alarm and
outflow pipe adapter
Liquid-outflow alarm and
inflow pipe adapter