7
6.
Up Button
: Increases light intensity in Run mode.
7.
Cable Clamp
: Grasps the light-source end of an inserted fiberoptic
cable. (When no cable is inserted, the bulb automatically turns off to
save bulb hours.)
8.
Jaw Handle
: Opens the fiberoptic-cable holder.
9.
Down Button
: Decreases light intensity in Run mode.
Figure 2
Rear panel of the X8000 console
10.
Voice-Control Port
: Connects to Stryker voice-control systems.
11.
SFB Series Connectors
: Enables FireWire connection with Stryker
FireWire devices. Provides connection for remote diagnoses and future
software upgrades.
12.
AC Inlet
: Connects to the provided power cord for AC power supply.
13.
Fuse Holder
: Contains two T 5.0 AH 250 V fuses.
14.
Equipotentiality Plug
: Provides a grounding post for common
grounding with other equipment. Connects to a potential equalization
conductor. The resulting medical electrical system shall follow all
applicable IEC 60601-1 requirements.
Stryker Endoscopy
5900 Optical Court
San Jose, CA 95138 USA
U.S. Patents:
5850496, 6110107, 6689050
Other patents pending
MADE IN USA