Electromagnetic Compatibility
Like other electrical medical equipment, the X8000 requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To
ensure electromagnetic compatibility (EMC), the X8000 must be installed and
operated according to the EMC information provided in this manual.
Note
The X8000 has been designed and tested to comply with IEC
60601-1-2 requirements for EMC with other devices.
Caution
Portable equipment may affect the normal function of the
X8000.
W
ARNING
D
O
NOT
USE
CABLES
OR
ACCESSORIES
OTHER
THAN
THOSE
PROVIDED
WITH
THE
X8000,
AS
THIS
MAY
RESULT
IN
INCREASED
ELECTROMAGNETIC
EMISSIONS
OR
DECREASED
IMMUNITY
TO
SUCH
EMISSIONS
.
W
ARNING
I
F
THE
X8000
IS
USED
ADJACENT
TO
OR
STACKED
WITH
OTHER
EQUIPMENT
,
OBSERVE
AND
VERIFY
NORMAL
OPERATION
OF
THE
X8000
IN
THE
CONFIGURATION
IN
WHICH
IT
WILL
BE
USED
PRIOR
TO
USING
IT
IN
A
SURGICAL
PROCEDURE
. C
ONSULT
THE
TABLES
BELOW
FOR
GUIDANCE
IN
PLACING
THE
X8000.
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
X8000 is intended for use in the electromagnetic environment specified below. The customer or the
user of X8000 should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment - guidance
RF emissions CISPR
11
Class B
X8000 is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage Fluctuations/
flicker emissions
IEC61000-3-3
Complies