20
4.
Insert the new bulb module (Stryker part number 220-201-000) along
the guide rails until it is fully seated on the mating connectors.
5.
Rotate the handle to the left to secure the bulb module within the light
source.
6.
Close and latch the bulb door.
7.
Power on the unit.
8.
Power on the bulb (if not already on) and verify that the LCD displays
the bulb hours as “0”.
Note
Using a non-Stryker replacement bulb module will result in no
light output.
Caution
Do not operate the X8000 with a burned-out bulb or with no
bulb installed.
Replacing the Fuses
1.
Unplug the light source from the AC outlet and remove the power cord
from the rear of the unit.
2.
Unlatch the fuse holder and remove the fuse(s).
3.
Replace the fuse(s) with fuses of the same value and rating.
W
ARNING
T
O
HELP
AVOID
THE
RISK
OF
FIRE
,
USE
ONLY
FUSES
RATED
AS
SPECIFIED
ON
THE
CONSOLE
.
4.
Reinstall the fuse holder.
Disposing of the X8000
The device must be disposed of according to local laws and hospital practices.
The device does not contain any hazardous materials.
This product is considered electronic equipment. It must not be disposed of as
unsorted municipal waste and must be collected separately. Please contact the
manufacturer or other authorized disposal company to decommission your
equipment.