Component Implantation
>
If needed, prepare the resected bone surfaces using
the Bone File, which is attached to the Impaction
Handle.
49
PS or CR Femoral Component -
Cemented/Cementless
>
Attach the Femoral Impactor Extractor to the
Impaction Handle and attach to the appropriate size
and side Femoral Component
(See Assembly 5)
.
Place the Femoral Component on the femur and
impact it until fully seated.
• Posterior Stabilized Knee: If Modular Femoral
Distal Fixation Pegs are to be used, assemble the
pegs to the Femoral Component using the 1/8
"
Hex Drive and the Slip Torque Handle prior to
implantation
(See Assembly 10)
.
>
The Femoral Impactor can be attached to the
Impaction Handle to further seat the Femoral
Component onto the prepared femur.
Note:
Clear all excess bone cement (Does not apply to
cementless component).