Patella Preparation
>
Determine the total thickness of the patella by using
the Patella Caliper.
>
There are two options for the patella preparation:
Bone Removing Method and Bone Remaining
Method.
45
Option 1
– Bone Removing Method
>
Assemble Patella Clamp Jaws to the Patella Clamp.
Attach the Patella Stylus to the circular hole on the top
side of either jaw by squeezing the gold tab
(See Assembly 9)
.
>
The Patella Stylus may swivel in this position to sweep
over the highest portion of the articular surface.
>
The Patella Stylus references the articular surface of
the patella in order to determine how much bone to
remove.
>
Set the desired resection amount on the Patella Stylus
by pressing the gold button and moving the body of
the Patella Stylus to the resection line.
>
Close the Patella Clamp around the patella.
>
The resection level should be set to match the
thickness of the appropriate size patella implant.
>
Ensure that the Patella Stylus is touching the desired
point(s) on the articular surface of the patella.
>
Make resection through one of the resection slots.