Figure 5
Figure 6
>
The Adjustment Block allows for a 8mm (the distal
thickness of the femoral component) and 10mm
(used to aid in the correction of a flexion contracture)
resection level.
>
Press the black button on the end of the Adjustment
Block and pull to set the resection to the desired level.
19
>
The Triathlon Knee System Instruments allow for a
clear view of the bone that is being resected to ensure
the appropriate level is set.
>
Slide the Adjustment Block Assembly within the
Femoral Alignment Guide until the Distal Resection
Guide contacts the anterior surface of the femur.