Figure 64
Instrument Bar
48
Figure 65
Patella Trial Assessment
>
Remove any residual cartilage and wash away all debris. Place correct
size Patella Trial (Symmetric or Asymmetric) onto the prepared patella.
>
Replace all Trials and assess patellar tracking by taking the knee through
a ROM. The patella should track normally throughout the ROM
without tendency for tilting or lateral subluxation.
27mm -
6541-3-627
29mm -
6541-3-629
31mm -
6541-3-631
33mm -
6541-3-633
36mm -
6541-3-636
39mm -
6541-3-639
Symmetric Patella Drill Template
29mm -
6541-3-617
32mm -
6541-3-618
35mm -
6541-3-619
38mm -
6541-3-620
40mm -
6541-3-621
Assymetric Patella Drill Template
6541-3-600
Patella Clamp
6541-3-801
Patella Clamp Base
6541-3-524
All-Poly Patella Drill w/Stop
6541-4-801
Universal Driver
P
a
te
ll
a
P
re
p
a
ra
ti
o
n
See Catalog
Symmetric Patella Trial
See Catalog
Asymmetric Patella Trial
>
Attach the All-Poly Patella Drill with Stop or Metal-Backed Patella Drill
for cementless metal-backed patella option to the Universal Driver and
drill through each fixation peg hole of the Patella Drill Template.
>
Disengage the Patella Clamp by pressing the release trigger. Press the
gold buttons on the Patella Clamp to remove the Patella Template.
6541-3-522
Metal-Backed Patella Drill