56
EN
Replacing Components and Modules
6.12
Uninstalling/Installing HPL Board
In case of malfunction of the gas tube heating and/or damage to the tube con-
nection, the HPL board must be replaced.
DANGER!
Disconnect device from power supply.
1. Open the device as described in chapter 3.6 Opening the Device .
2. Remove device front panel as described in chapter 6.14 Uninstalling/Install-
3. Remove the ST1 plug /cable
(2)
from the rear HPL board
(1)
by pulling paral-
lel to the top.
4. Unscrew the two fastening screws
(3)
and remove the HPL board. Make sure
the O-ring remains in place when removing the component.
5. Follow these steps in reverse for assembly. Make sure the O-ring is in place
when installing the component.
The following tests must be performed after installing a new HPL board:
Fig. 6-27 Position of HPL board
(1)
HPL board
(2)
ST1 plug/flat ribbon cable
(3)
2x fastening screws M2.5
(3)
(1)
(2)
(3)
Fig. 6-28 Position of HPL board
(1)
HPL board
(2)
ST1 plug/flat ribbon cable
(3)
2x fastening screws M2.5
(3)
(1)
(2)
(3)
Содержание PneumoSure