58
EN
Replacing Components and Modules
6.14
Uninstalling/Installing Device Front
DANGER!
Disconnect device from power supply.
1. Open the device as described in chapter 3.6 Opening the Device .
2. Remove the three fastening screws
(8)
on the bottom of the device (under-
side).
3. Remove the two cap nuts
(3)
of the con-rods on rear of device.
4. Turn the two con-rods
(2)
counterclockwise to remove them from the front
of the device. This can be done with the help of a hex key or Allen wrench, for
example, inserted into the hole of the con-rod.
5. Remove fluid sensor from tube connection as described in chapter 6.13 Unin-
stalling/Installing Fluid Sensor .
6. Remove the patient pressure measurement tube at position
(6)
.
7. Remove the plug ST4/cable leading to the power switch LED from the NET
Fig. 6-30 Position of device front
(1)
Device front
(2)
2x con-rods
(3)
2x cap nuts (device rear)
(4)
ST8 plug on MIO board/cable lead-
ing to PCU
(5)
Fluid sensor
(6)
Patient pressure measurement
connection
(1)
(2)
(5)
(4)
(3)
(6)
Fig. 6-31 Device base
(7)
Device front
(8)
3x fastening screws
(7)
(8)