Maintenance and Service Tasks
21
EN
4.10
Testing the Occlusion Warning
1. In high flow mode, set the nominal pressure = 15 mm Hg and the nominal
flow = 3 l/min (see also 4.9 Testing the Venting System and Overpressure
Alarm).
2. Attach the calibration adapter
(3)
to the device as described in chapter 4.7
Mounting the Calibration Adapter and turn the 3-port valve
(2)
to the depict-
ed position.
3. Start insufflation with the START/STOP key.
4. An audible warning signal is emitted and
Occlusion
is depicted (note: The
acoustic warning signal can be deactivated in the user menu!)
5. Turn the 3-port valve to the right by 90°. The warnings are deactivated.
6. End insufflation with the START/STOP key.
Corrective measure
• Check (for leaks) to make sure the calibration adapter functions properly.
• Replace the LPU.
Fig. 4-13 Testing the occlusion warning
(1)
Tube, small diameter
(2)
Three-port valve
(3)
Calibration adapter
mm Hg
(2)
(3)
(1)