Reprocessed HARMONIC ACE

®

+7, 5mm Diameter Shears with  

Advanced Hemostasis  

Page 6 of 19 

 

 
 

 

To avoid user or patient injury, do not activate an electrosurgical device in close proximity to the HARMONIC instruments. 
The aerosols created by the activation of the HARMONIC instruments in fatty tissue are potentially flammable. 

 

Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad 

to avoid damage to the tissue pad and increased blade, clamp arm and distal shaft temperatures. 

 

The entire exposed blade tip and any exposed blade shaft are active and will cut/coagulate tissue when the instrument 

blade is activated. Be careful to avoid inadvertent contact between all exposed blade surfaces and surrounding tissue 
when using the instrument. 

 

Use only the appropriate Foot Switch, Hand Piece, instruments, and power cord to ensure that they are compatible with 

the generator. 

 

After removing the instrument, examine the tissue for hemostasis. If hemostasis is not present, appropriate techniques 

should be used to achieve hemostasis. 

 

Successful hemostasis may require adjunct measures when HARMONIC instruments are used on solid organs. Due to the 

difficulty of visualizing internal structures, proceed slowly and do not attempt to transect large masses of tissue in one 
activation. Avoid the division of large vascular/biliary bundles when using the instrument under these conditions. 

 

If activation is unintentionally stopped while sealing, maintain jaw closure and reactivate. 

 

Do not use Advanced Hemostasis mode for procedures where energy application is desired prior to full closure of the jaws 

(e.g. solid organs). Energy is not delivered in Advanced Hemostasis mode until the jaws are completely closed. 

 

During benchtop testing of vessels >5 mm, the strongest vessel seals were achieved by allowing the Advanced Hemostasis 

mode to completely transect the targeted vessel. 

 

Prolonged usage of Advanced Hemostasis Mode may cause tissue pad damage. 

 

Products manufactured or distributed by companies not authorized by Ethicon Endo-Surgery may not be compatible with 

the HARMONIC system. Use of such products may lead to unanticipated results and possible injury to the user or patient. 

 

If during use the generator displays ''Advanced Features Are Not Available In This Device", the following functions of 

Adaptive Tissue Technology are no longer available: Regulated Energy Delivery, Enhanced Audible Feedback and 
Advanced Hemostasis. As a consequence, the device vessel sealing indication does not exceed 5 mm. 

 

Instruments or devices which come into contact with bodily fluids may require special disposal handling to prevent 

biological contamination. 

 

Incidental and prolonged activation against solid surfaces, such as bone, may result in blade heating and subsequent blade 

failure, and should be avoided. 

 

Dispose of all opened instruments whether used or unused. 

 

This device is packaged and sterilized for single use only. Multiple patient use may compromise the device integrity or 

create a risk of contamination that, in turn, may result in patient injury or illness. 

 
Directions for Use 

1.

 

Verify compatibility of all instruments and accessories prior to using this instrument (refer to 

Warnings and 

Precautions

). 

2.

 

The Hand Piece is shipped non-sterile. It must be sterilized prior to each use according to the instructions for use supplied 
with the Hand Piece. 

 

Assembly 

1.

 

Using sterile technique, remove the instrument from the package. To avoid damage, do not flip the instrument into the 
sterile field. 

2.

 

While holding the Hand Piece in a vertical orientation, attach the Hand Piece to the instrument by rotating the instrument 
onto the Hand Piece in a clockwise rotation as viewed from the distal end of the instrument (finger tight only). 

3.

 

Use the torque wrench (already mounted to the shaft) to tighten the blade onto the Hand Piece. Turn the torque wrench 
clockwise while holding only the gray Hand Piece until it clicks twice indicating that sufficient torque has been applied to 

secure the blade. 

4.

 

Note: Do not use any other means than the torque wrench to attach or detach the instrument from the Hand Piece. 

5.

 

Note: Do not torque the instrument by hand without the torque wrench or damage may occur to the Hand Piece. 

6.

 

Note: Hold only the gray Hand Piece and not the instrument handle while applying the torque wrench (Illustration 2). 

7.

 

Close the trigger. Remove the torque wrench by sliding it off of the shaft. Do not dispose of the torque wrench until the 
procedure is completed. The torque wrench is used to remove the instrument from the Hand Piece following the 

procedure (Illustration 3). Dispose of the torque wrench only after completing the procedure. 

8.

 

Note: Take care to avoid damage to the blade and clamp arm by closing the trigger while sliding the torque wrench onto or 
off of the shaft. 

Содержание Harmonic Ace+7

Страница 1: ...se Caution Federal U S A law restricts this device to sale by or on the order of a physician STERILE Explanation of Symbols Sterilized by Ethylene Oxide Gas Date of Reprocessing Use by Date Product Code Do Not Reuse See Instructions For Use Stryker Sustainability Solutions Inc 2016 1810 W Drake Dr Tempe AZ 85283 sustainability stryker com 888 888 3433 ...

Страница 2: ...Reprocessed HARMONIC ACE 7 5mm Diameter Shears with Advanced Hemostasis Page 2 of 19 ...

Страница 3: ...Reprocessed HARMONIC ACE 7 5mm Diameter Shears with Advanced Hemostasis Page 3 of 19 ...

Страница 4: ... represent relative vessel size The MAX button is typically used for smaller vessels where cutting speed is fastest The MIN button is typically used in slightly larger vessels and has reduced cutting speed It is indicated for vessels up to 5 mm in size The Advanced Hemostasis button is designed for larger vessels and is indicated for vessels up to 7 mm in size In this mode cutting speed is further...

Страница 5: ...em failure ensure the availability of the appropriate back up equipment relevant to the specific procedure Audible high pitched ringing resonating from the blade or Hand Piece are an abnormal condition and an indicator that the blade or Hand Piece is not operating properly The ringing may be an indicator that the Hand Piece is beyond its useful life or that the blade has not been attached properly...

Страница 6: ... the following functions of Adaptive Tissue Technology are no longer available Regulated Energy Delivery Enhanced Audible Feedback and Advanced Hemostasis As a consequence the device vessel sealing indication does not exceed 5 mm Instruments or devices which come into contact with bodily fluids may require special disposal handling to prevent biological contamination Incidental and prolonged activ...

Страница 7: ...tasis mode until the jaws are completely closed Caution If activation is unintentionally stopped while sealing maintain jaw closure and reactivate Note The generator provides an audible tone to indicate when the instrument blade is active The generator changes to a second activation tone as Adaptive Tissue Technology regulates the delivery of energy Thermal influences such as fluids or minimal to ...

Страница 8: ...ucts Stryker warrants all reprocessed products subject to the exceptions provided herein to be free from defects in reprocessing and to substantially conform to the product specifications contained in the documentation provided by Stryker with the products for one use in accordance with the instructions for use of such product STRYKER SHALL NOT BE LIABLE FOR ANY DAMAGES TO THE EXTENT CAUSED BY ANY...

Страница 9: ...ker determines in its reasonable discretion that the claimed defect or non conformance in the product is excluded from warranty coverage as described hereunder it will notify the customer of such determination and will provide an estimate of the cost of repair of the product In such an event any repair would be performed at Stryker s standard rates Products and product components repaired or repla...

Страница 10: ...édérale américaine É U ce dispositif ne peut être vendu que par un médecin ou sur ordonnance médicale STÉRILE Explication des symboles Stérilisé par oxyde d éthylène Date du retraitement Date de péremption Code du produit Ne pas réutiliser Voir le mode d emploi Stryker Sustainability Solutions Inc 2016 1810 W Drake Dr Tempe AZ 85283 États Unis sustainability stryker com 888 888 3433 ...

Страница 11: ...Cisailles HARMONIC ACE 7 retraitées de 5 mm de diamètre pour hémostase avancée Page 11 de 19 ...

Страница 12: ...Cisailles HARMONIC ACE 7 retraitées de 5 mm de diamètre pour hémostase avancée Page 12 de 19 ...

Страница 13: ...ent peuvent pivoter sur 360 pour faciliter la visualisation et l accès aux tissus ciblés Les trois tirets affichés sur l instrument représentent une dimension relative du vaisseau Le bouton MAX est généralement utilisé pour les vaisseaux plus petits lorsque la vitesse de la coupe est plus rapide Le bouton MIN est généralement utilisé pour les vaisseaux légèrement plus grands et que la vitesse de c...

Страница 14: ...ntion chirrugicale Une compréhension approfondie des principes et techniques impliqués dans les procédures laser électrochirurgicales et ultrasoniques est essentielle pour éviter tout risque d électrocution et de brûlure infligée à la fois au patient et au personnel médical ainsi que pour éviter tout dommage au dispositif ou aux autres instruments médicaux Assurez vous que ni l isolation électriqu...

Страница 15: ...ne fois l instrument retiré examinez les tissus pour déceler une mauvaise hémostase Si l hémostase est inadéquate utilisez les techniques appropriées pour obtenir l hémostase Une hémostase fructueuse peut demander des mesures supplémentaires lorsque les instruments HARMONIC sont utilisés sur des organes pleins En raison de la difficulté de visualisation des structures internes procédez lentement e...

Страница 16: ...niveau de puissance minimal voulu à l aide des boutons INCREASE augmenter et DECREASE diminuer qui apparaissent à l écran du générateur Remarque Le niveau de puissance minimal de démarrage recommandé est 3 Illustration 4 Pour une vitesse accrue de coupe des tissus utilisez un niveau de puissance plus élevé du générateur et pour une coagulation accrue utilisez plutôt un niveau de puissance plus fai...

Страница 17: ...te du déclencheur Remarque Maintenez le bras de serrage ouvert lorsque vous utilisez la partie inférieure interne de la lame pour une coupe inverse Illustration 10 AVERTISSEMENT N utilisez pas le mode d hémostase avancée pour les procédures où une application d énergie est requise avant la fermeture complète des mâchoires p ex dans le cas des organes pleins En mode d hémostase avancée l énergie es...

Страница 18: ...AUCUN CAS LA RESPONSABILITÉ DE STRYKER EN LIEN AVEC LA VENTE DU PRODUIT QUE CE SOIT DANS LE CADRE DES THÉORIES DE LA RUPTURE D UN CONTRAT D UN TORT D UNE FAUSSE REPRÉSENTATION D UNE FRAUDE D UNE GARANTIE D UNE NÉGLIGENCE D UNE RESPONSABILITÉ STRICTE OU DE TOUTE AUTRE THÉORIE DU DROIT NE PEUT ÊTRE SUPÉRIEURE AU PRIX D ACHAT À LA VALEUR ACTUELLE DU MARCHÉ OU À LA VALEUR RÉSIDUELLE DES PRODUITS SELON...

Страница 19: ...mplacé la propriété de Stryker Si un remboursement est fourni par Stryker le produit pour lequel le remboursement est décidé doit être retourné à Stryker et deviendra la propriété de Stryker HARMONIC et HARMONIC ACE sont des marques de commerce d Ethicon Endo Surgery ULS EL10053 Rév E 10 2017 RM702506 Les informations concernant le fabricant d équipement d origine fournies sur l étiquette apparais...

Отзывы: