Stryker Harmonic Ace+7 Скачать руководство пользователя страница 5

Reprocessed HARMONIC ACE

®

+7, 5mm Diameter Shears with  

Advanced Hemostasis  

Page 5 of 19 

 

 
 

Indications for Use 

The Reprocessed HARMONIC ACE

®

+7, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions 

when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for 

electrosurgery, lasers and steel scalpels in general, plastic, pediatric, gynecologic, urologic, thoracic, exposure to orthopedic  
structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic 
procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced 
Hemostasis hand control button. 

 

Contraindications for Use 

Reprocessed HARMONIC ACE

®

+7, 5 mm Diameter Shears with Advanced Hemostasis are contraindicated for:   

 

The instruments are not indicated for incising bone. 

 

The instruments are not intended for contraceptive tubal occlusion. 

 

Warnings and Precautions 

 

Caution: Federal (USA) law restricts this device to sale by or on the order of a physician. 

 

Minimally invasive procedures should be performed only by persons having adequate training and familiarity with 

minimally invasive techniques. Consult medical literature relative to techniques, complications, and hazards prior to 
performance of any minimally invasive procedure. 

 

Minimally invasive instruments may vary in diameter from manufacturer to manufacturer. When minimally invasive 

instruments and accessories from different manufacturers are employed together in a procedure, verify compatibility 
prior to initiation of the procedure. 

 

A thorough understanding of the principles and techniques involved in laser, electrosurgical, and ultrasonic procedures is 
essential to avoid shock and burn hazards to both patient and medical personnel and damage to the device or other 

medical instruments. Ensure that electrical insulation or grounding is not compromised. Do not immerse instruments in 
liquid unless the instruments are designed and labeled to be immersed. 

 

Verify compatibility with generators. Use device only with Ethicon Endo-Surgery Generator G11 (GEN11) software 

version 2013_1, 2014_1, or 2016_1. Software revision can be found under "System Information" in the Generator G11 
(GEN11) "Settings" menu. Refer to the Generator G11 (GEN11) Operator's Manual for more information. 

 

In case of system failure, ensure the availability of the appropriate back up equipment relevant to the specific procedure. 

 

Audible high-pitched ringing, resonating from the blade or Hand Piece, are an abnormal condition and an indicator that 

the blade or Hand Piece is not operating properly. The ringing may be an indicator that the Hand Piece is beyond its useful 
life or that the blade has not been attached properly, which may result in abnormally high shaft temperatures and user or 

patient injury. 

 

The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced 

Hemostasis hand control button. Do not attempt to seal vessels in excess of 7 mm in diameter. 

 

Blood and tissue buildup between the blade and shaft may result in abnormally high temperatures at the distal end of the 

shaft. To prevent burn injury, remove any visible tissue buildup at the distal end of the shaft. 

 

As with all energy sources (Electrosurgery, Laser, or Ultrasound), there are concerns about the carcinogenic and infectious 

potential of the by-products, such as tissue smoke plume and aerosols. Appropriate measures such as protective eyewear, 
filtration masks, and effective smoke evacuation equipment should be used in both open and laparoscopic procedures. 

 

Do not attempt to bend, sharpen, or otherwise alter the shape of the blade. Doing so may cause blade failure and user or 

patient injury. 

 

To avoid user or patient injury in the event that accidental activation occurs, the instrument blade, clamp arm, and distal 

end of the shaft should not be in contact with the patient, drapes, or flammable materials while not in use. 

 

During and following activation in tissue, the instrument blade, clamp arm, and distal 7 cm of the shaft may be hot. Avoid 

unintended contact with tissue, drapes, and surgical gowns at all times. 

 

Avoid contact with any and all other instruments while the instrument is activated. Contact with staples, clips or other 

instruments while the instrument is activated may result in cracked or broken blades. 

 

Do not introduce or withdraw the instrument with the jaws open through a trocar sleeve as this may damage the 

instrument. 

 

Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between 

them. Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher 
blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument. If this occurs, there 

may be an instrument failure, and the generator touchscreen displays a troubleshooting message. 
 

Содержание Harmonic Ace+7

Страница 1: ...se Caution Federal U S A law restricts this device to sale by or on the order of a physician STERILE Explanation of Symbols Sterilized by Ethylene Oxide Gas Date of Reprocessing Use by Date Product Code Do Not Reuse See Instructions For Use Stryker Sustainability Solutions Inc 2016 1810 W Drake Dr Tempe AZ 85283 sustainability stryker com 888 888 3433 ...

Страница 2: ...Reprocessed HARMONIC ACE 7 5mm Diameter Shears with Advanced Hemostasis Page 2 of 19 ...

Страница 3: ...Reprocessed HARMONIC ACE 7 5mm Diameter Shears with Advanced Hemostasis Page 3 of 19 ...

Страница 4: ... represent relative vessel size The MAX button is typically used for smaller vessels where cutting speed is fastest The MIN button is typically used in slightly larger vessels and has reduced cutting speed It is indicated for vessels up to 5 mm in size The Advanced Hemostasis button is designed for larger vessels and is indicated for vessels up to 7 mm in size In this mode cutting speed is further...

Страница 5: ...em failure ensure the availability of the appropriate back up equipment relevant to the specific procedure Audible high pitched ringing resonating from the blade or Hand Piece are an abnormal condition and an indicator that the blade or Hand Piece is not operating properly The ringing may be an indicator that the Hand Piece is beyond its useful life or that the blade has not been attached properly...

Страница 6: ... the following functions of Adaptive Tissue Technology are no longer available Regulated Energy Delivery Enhanced Audible Feedback and Advanced Hemostasis As a consequence the device vessel sealing indication does not exceed 5 mm Instruments or devices which come into contact with bodily fluids may require special disposal handling to prevent biological contamination Incidental and prolonged activ...

Страница 7: ...tasis mode until the jaws are completely closed Caution If activation is unintentionally stopped while sealing maintain jaw closure and reactivate Note The generator provides an audible tone to indicate when the instrument blade is active The generator changes to a second activation tone as Adaptive Tissue Technology regulates the delivery of energy Thermal influences such as fluids or minimal to ...

Страница 8: ...ucts Stryker warrants all reprocessed products subject to the exceptions provided herein to be free from defects in reprocessing and to substantially conform to the product specifications contained in the documentation provided by Stryker with the products for one use in accordance with the instructions for use of such product STRYKER SHALL NOT BE LIABLE FOR ANY DAMAGES TO THE EXTENT CAUSED BY ANY...

Страница 9: ...ker determines in its reasonable discretion that the claimed defect or non conformance in the product is excluded from warranty coverage as described hereunder it will notify the customer of such determination and will provide an estimate of the cost of repair of the product In such an event any repair would be performed at Stryker s standard rates Products and product components repaired or repla...

Страница 10: ...édérale américaine É U ce dispositif ne peut être vendu que par un médecin ou sur ordonnance médicale STÉRILE Explication des symboles Stérilisé par oxyde d éthylène Date du retraitement Date de péremption Code du produit Ne pas réutiliser Voir le mode d emploi Stryker Sustainability Solutions Inc 2016 1810 W Drake Dr Tempe AZ 85283 États Unis sustainability stryker com 888 888 3433 ...

Страница 11: ...Cisailles HARMONIC ACE 7 retraitées de 5 mm de diamètre pour hémostase avancée Page 11 de 19 ...

Страница 12: ...Cisailles HARMONIC ACE 7 retraitées de 5 mm de diamètre pour hémostase avancée Page 12 de 19 ...

Страница 13: ...ent peuvent pivoter sur 360 pour faciliter la visualisation et l accès aux tissus ciblés Les trois tirets affichés sur l instrument représentent une dimension relative du vaisseau Le bouton MAX est généralement utilisé pour les vaisseaux plus petits lorsque la vitesse de la coupe est plus rapide Le bouton MIN est généralement utilisé pour les vaisseaux légèrement plus grands et que la vitesse de c...

Страница 14: ...ntion chirrugicale Une compréhension approfondie des principes et techniques impliqués dans les procédures laser électrochirurgicales et ultrasoniques est essentielle pour éviter tout risque d électrocution et de brûlure infligée à la fois au patient et au personnel médical ainsi que pour éviter tout dommage au dispositif ou aux autres instruments médicaux Assurez vous que ni l isolation électriqu...

Страница 15: ...ne fois l instrument retiré examinez les tissus pour déceler une mauvaise hémostase Si l hémostase est inadéquate utilisez les techniques appropriées pour obtenir l hémostase Une hémostase fructueuse peut demander des mesures supplémentaires lorsque les instruments HARMONIC sont utilisés sur des organes pleins En raison de la difficulté de visualisation des structures internes procédez lentement e...

Страница 16: ...niveau de puissance minimal voulu à l aide des boutons INCREASE augmenter et DECREASE diminuer qui apparaissent à l écran du générateur Remarque Le niveau de puissance minimal de démarrage recommandé est 3 Illustration 4 Pour une vitesse accrue de coupe des tissus utilisez un niveau de puissance plus élevé du générateur et pour une coagulation accrue utilisez plutôt un niveau de puissance plus fai...

Страница 17: ...te du déclencheur Remarque Maintenez le bras de serrage ouvert lorsque vous utilisez la partie inférieure interne de la lame pour une coupe inverse Illustration 10 AVERTISSEMENT N utilisez pas le mode d hémostase avancée pour les procédures où une application d énergie est requise avant la fermeture complète des mâchoires p ex dans le cas des organes pleins En mode d hémostase avancée l énergie es...

Страница 18: ...AUCUN CAS LA RESPONSABILITÉ DE STRYKER EN LIEN AVEC LA VENTE DU PRODUIT QUE CE SOIT DANS LE CADRE DES THÉORIES DE LA RUPTURE D UN CONTRAT D UN TORT D UNE FAUSSE REPRÉSENTATION D UNE FRAUDE D UNE GARANTIE D UNE NÉGLIGENCE D UNE RESPONSABILITÉ STRICTE OU DE TOUTE AUTRE THÉORIE DU DROIT NE PEUT ÊTRE SUPÉRIEURE AU PRIX D ACHAT À LA VALEUR ACTUELLE DU MARCHÉ OU À LA VALEUR RÉSIDUELLE DES PRODUITS SELON...

Страница 19: ...mplacé la propriété de Stryker Si un remboursement est fourni par Stryker le produit pour lequel le remboursement est décidé doit être retourné à Stryker et deviendra la propriété de Stryker HARMONIC et HARMONIC ACE sont des marques de commerce d Ethicon Endo Surgery ULS EL10053 Rév E 10 2017 RM702506 Les informations concernant le fabricant d équipement d origine fournies sur l étiquette apparais...

Отзывы: