![Stryker EOLE DC 32](http://html1.mh-extra.com/html/stryker/eole-dc-32/eole-dc-32_service-manual_1385708005.webp)
www.stryker.com
836004-5210 V1.1
3
Service
PUMPING SET REPLACEMENT
TOOLS REQUIRED:
●
Phillips screwdriver for M3 screws
●
Threadlocker (LOCTITE 290)
PROCEDURE:
1.
Take out the 8 rubbers and remove 8 screws on the rear case. Take off the rear case.
2.
Remove 6 screws on the base.
3.
Remove wire connections to power switch and front panel PCBA. Take out the module set from the
case.
4.
Pumping set is inside the compressor. Unscrew the 4 screws on the compressor and open the case.
5.
Remove the screw of pumping set.
6.
Take out the pumping set.