Problem/Failure

Possible Cause

Recommended Action 

Low pressure alarm

Pump performance degradation

Replace pumping set 
Replace filter

CPR control is not fully closed

Close CPR control

There is a leak in the mattress

Replace broken tube

The tubing of mattress is 
not connected properly

Check the tubing connectors and make 
sure they are securely connected

Tubing inside the pump is 
not connected properly

Open the pump housing and connect 
the internal tubing properly.

Power failure alarm

Fuse is burnt

Replace fuse

No power 

Make sure the unit is plugged into ground outlet

Transformer fail

Replace main PCBA

Main PCBA fail

Replace main PCBA.

Service alarm

Main PCBA fail

Replace main PCBA

Timing motor fail

Replace timing motor module

Micro switch fail

Replace micro switch

Panel display error

Button or LED fail

Replace front panel PCBA

Cannot stop buzzer beeping

Forget to switch off power button

Switch off power switch

The buzzer wire is not connected 
to the correct position

Check the wire connections on front 
panel and on main PCBA

Troubleshooting

Quick Reference Replacement Parts

Part Name

Part Number

For product model

Pumping set (50Hz, 8 Liter)

2870001016

EOLE DC/EOLE

Timing motor module

2870001017

EOLE DC/EOLE

Compressor

2870001004

EOLE DC/EOLE

Micro switch

2870001018

EOLE DC/EOLE

Main PCBA

2871001002

EOLE DC

Main PCBA

2870001014

EOLE

Front panel PCBA

2870001015

EOLE DC/EOLE

Filter

2870001019

EOLE DC/EOLE

Fuse (T1AL/250V)

2870001002

EOLE DC/EOLE

Label, Stryker, front enclosure

2870001006

EOLE DC/EOLE

Screw cover, rubber

2870001007

EOLE DC/EOLE

Screw, M3 x 20 mm

2870001008

EOLE DC/EOLE

Screw, M3 x 10 mm

2870001011

EOLE DC/EOLE

Air outlet assembly, with CPC connector

2870001009

EOLE DC/EOLE

Switch assembly, with PVC switch cap

2870001010

EOLE DC/EOLE

Front case enclosure

2870001012

EOLE DC/EOLE

Mattress bottom cover 32” (80cm)

2871019009

EOLE DC 

Back Case

2870001013

EOLE/EOLE DC

Pump Service Manual for Eole and Eole DC

2870009002

EOLE/EOLE DC

Содержание EOLE DC 32"

Страница 1: ...2017 10 A 0 836004 5210 V1 1 www stryker com Service Manual EOLE EOLE DC Pump REF 2870 REF 2871...

Страница 2: ......

Страница 3: ...ment Parts 2 Service 3 Pumping set replacement 3 Air cell replacement 5 Check internal tubing 5 Fuse replacement 6 Main PCBA replacement 6 Timing motor module replacement 8 Micro switch replacement 9...

Страница 4: ...ing Forget to switch off power button Switch off power switch The buzzer wire is not connected to the correct position Check the wire connections on front panel and on main PCBA Troubleshooting Quick...

Страница 5: ...the 8 rubbers and remove 8 screws on the rear case Take off the rear case 2 Remove 6 screws on the base 3 Remove wire connections to power switch and front panel PCBA Take out the module set from the...

Страница 6: ...pumping set and hold on this position and then tighten the screw 2 Press the vibration rods to make sure they will not touch the front object 3 Apply threadlocker around the screw to set for the best...

Страница 7: ...air cells there is an air leakage If there is a leakage replace the air cell CHECK INTERNAL TUBING TOOLS REQUIRED Phillips screwdriver for M3 screws PROCEDURE 1 Remove the 8 rubber stops Using a Phili...

Страница 8: ...Remove the rear case 2 Push the fuse holder together with turning counterclockwise to open it 3 Replace the fuse and discard the old one MAIN PCBA REPLACEMENT TOOLS REQUIRED Phillips screwdriver for M...

Страница 9: ...e connections on the main PCBA are as below 7 After replacement perform pressure calibartion Refer to Pressure Calibration section 4 Remove the 2 screws and all wire connections on the main PCBA 3 Rem...

Страница 10: ...ber stops Using a Philips screwdriver remove the 8 screws that hold the rear case Remove the rear case 2 Remove the tubes and wire connections from timing motor module 3 Remove the 2 screws on the mot...

Страница 11: ...screws PROCEDURE 1 Remove the 8 rubber stops Using a Philips screwdriver remove the 8 screws that hold the rear case Remove the rear case 2 Remove the wire connections from micro switch 3 Remove the...

Страница 12: ...the 8 rubber stops Using a Philips screwdriver remove the 8 screws that hold the rear case Remove the rear case 2 Remove the tubes and wire connections from front panel PCBA 3 Remove the 5 screws on...

Страница 13: ...bers and unscrew 8 screws on the rear case Remove the rear case 2 Connect the deivce to power supply and switch on power switch 3 In order press A B C A then it goes into calibration mode 4 For timing...

Страница 14: ...press A There will be a long beep 11 Push or pull the loading piece of the syringe until the pressure meter shows and holds 63 mmHg and then press A There will be a long beep After about 5 seconds ano...

Страница 15: ...OCEDURE 1 Connect the pressure test tool to the pump Photo is for reference only 2 Switch on power 3 After inflation is completed change mode to Static 4 Choose leve1 and after the pressure is stable...

Страница 16: ...2017 10 836004 5210 V1 1 www stryker com Stryker Medical 3800 E Centre Avenue Portage Michigan 49002 USA Stryker European Operations B V Herikerbergweg 110 Amsterdam 1101 CM Netherlands...

Отзывы: