PRESSURE TEST
TOOLS REQUIRED:
●
Air cells with air loss x2
●
Pressure meter
●
PVC tubes and connector.
Note:
Pressure Meter suggested
Specification
±2psi or ±5psi range
Resolution: 0.001psi
Accuracy: ±0.3% FS
PROCEDURE:
1.
Connect the pressure test tool to the pump.
Photo
is
for
reference
only
2.
Switch on power
3.
After inflation is completed, change mode to “Static”.
4.
Choose leve1, and after the pressure is stable, check the pressure meter. The pressure should be
20 ±
3 mmHg
.
5.
Choose level 5, and after the pressure is stable, check the pressure meter. The pressure should be
32
± 5 mmHg
.
Note:
If the pressure is out of range, please repeat the calibration process.