0986-001-011 Rev-R
EN
www.stryker.com
5
CastVac Wall Mount Bracket
NOTES:
▪
The CastVac Wall Mount Bracket is supplied with CastVac REF 0987-000-
000 only.
▪
For installation instructions and more information, see the instructions for
use supplied with the CastVac Wall Mount Bracket.
AB
AA
AA
Cast Cutter Hook (2)
AB
Tool Bracket
Function Switch
Standby Mode –
When a Cast Cutter is connected to
the power outlet, CastVac operation is controlled by the
Cast Cutter function switch. The CastVac will operate
only when the Cast Cutter function switch is set to the
standard or high speed mode.
NOTE:
In standby mode, CastVac REF 0996-000-000
will continue to operate for 10 seconds after the Cast
Cutter is turned off.
Off –
The CastVac will not operate.
On –
The CastVac will operate continuously.
Instructions
To Connect the CastVac to a Cast Cutter and
a Power Source
NOTE:
To provide optimum power, the vacuum hose must be securely
connected to both the hose connector and the Cast Cutter.
1. Depress the button on the hose cuff and align it with the hole in the hose
connector.
2. Push the hose into the hose connector and release the button. Make sure
the button is completely positioned and seated in the hole.
3. Remove the vacuum line cap from the Cast Cutter.
4. Depress the button on the hose cuff and align it with the hole in the Cast
Cutter.
5. Push the hose into the Cast Cutter and release the button. Make sure the
button is completely positioned and seated in the hole.