44
sterilization tray, consult the instructions provided with the tray for proper
sterilization parameters. Stryker recommends sterilizing the device inside
of a sterilization tray.
•
Sterilize only one camera head per tray, or incomplete sterilization may
result. Follow any instructions provided with the sterilization tray or system
regarding tray setup and other devices that may be sterilized within the
same tray.
•
Devices repaired by or purchased from third-party service organizations
could expose patients to significant risk. These devices are no longer
validated by Stryker for cleanliness, disinfection, and sterilization, or for
safety and efficacy.
•
The user shall defer to the facility’s procedures regarding occupational
exposure to bloodborne pathogens.
Cautions
•
Always install the soaking cap prior to processing the camera. Failure to
properly tighten the soaking cap will corrode the connector pins and void
the warranty. Refer to the Installing the Soaking Cap section for more detail
about installing the cap.
•
Inspect the camera cable for cuts and breaks before soaking in any fluid.
Return any damaged camera to Stryker for service.
•
Never store the camera in the same tray with sharp instruments. Do not
soak the camera while it is inside a tray.
•
Do not use brushes or pads with metal or abrasive tips during manual
cleaning, as permanent scoring or damage could result.
•
To minimize galvanic corrosion, avoid soaking dissimilar metals in close
proximity.
•
The device cannot withstand an automated disinfection method.
•
The 1688 camera heads are not autoclavable. Steam sterilizing camera
heads that are not marked
AUTOCLAVE
will result in product damage.
•
Allow the camera head to cool before connecting it to the console.
Connecting the camera head while it is hot may result in system error.
Limitations on Reprocessing
•
Do not cross-sterilize the device. Using multiple sterilization methods may
significantly reduce the performance of the device.
•
Repeated automated cleaning can degrade the product’s cosmetic
appearance.
•
Damage caused by improper processing is not covered by the warranty.