36
Programming Camera Head Buttons
The camera head buttons can be customized differently for each surgical
specialty.
Contact a Stryker representative for assistance with button
programming.
The button configuration for the selected surgical speciality will appear on
the display monitor when the camera head is connected to the console. The
button configuration will disappear once the Camera button is pressed.
Note:
The default up and down camera head button functions when SPY
mode is on cannot be reconfigured.
The following functions can be programmed to the buttons:
Function Name
Function Description
AUTO LIGHT TOGGLE Toggle the Auto Light function on/off
BACKLIGHT UP
Increase Backlight level, which affects the brightness
of surrounding anatomy in the camera image that is
not displayed as fluorescing green
BACKLIGHT DOWN
Decrease Backlight level
BACKLIGHT CYCLE
Increase Backlight level until maximum level, then
cycle back to minimum level
BRIGHTNESS UP
Increase brightness level
BRIGHTNESS DOWN
Decrease brightness level
BRIGHTNESS CYCLE
Increase brightness until maximum level, then cycle
back to minimum level
ENHANCE UP
Increase enhance level, which sharpens the camera
image
ENHANCE DOWN
Decrease enhance level
ENHANCE CYCLE
Increase enhance until maximum level, then cycle
back to minimum level
HUB FUNCTION
Signal the device control console to perform an
assignable command
LIGHT SOURCE
TOGGLE
Toggle the light source between activating and
deactivating light output
MENU OPEN
Open the menu with device control options
NO FUNCTION
No function
PICTURE
Activate picture function on digital capture console
(activate Remote 1 cable)
Содержание 1688 Series
Страница 2: ......
Страница 22: ...18 Wiring Diagram 0240200100 1688010000 0220230300 0240 031 050 3 4 2 6 1 5 ...