41
Problem
Possible Solution
Optics are dirty
•
Rotate the scope. If dust particles in the
picture rotate, the dust is located on the
scope itself. Follow the manufacturer’s
instructions for cleaning the eyepiece and
negative lens.
•
If particles in the picture do not move
when you rotate the scope, the particles are
located on the coupler or camera. Remove
the scope and clean the window on the front
of the coupler with a dry or alcohol-tipped
cotton swab.
•
Disassemble the scope, coupler, and camera
head, and clean and dry all windows on the
components.
•
Ensure all components are completely dry
before reassembling them, or fogging may
result.
Blurry picture
•
Ensure the coupler or C-mount scope is in
focus.
•
On the Home screen, ensure the surgical
specialty is not set to FLEXI-SCOPE unless you
are using a flexible scope.
•
Disassemble the scope, coupler, and camera
head, and clean and dry all windows on the
components.
SIDNE device does not
recognize camera head
•
Contact your Stryker representative for
compatibility settings.
SPY mode won’t turn
on
•
Confirm the requirements in the SPY Mode
Requirements section have been met.
Visual artifacts
observed when Auto
Light is on
•
Turn off Auto Light from the camera, and
manually reduce the brightness from the
light source.
Note:
If this Troubleshooting section does not resolve the problem, call
Stryker Technical Support at 1-877-478-7953 (inside the U.S.) or refer to the
warranty.