Guidance and Manufacturer’s Declaration--Electromagnetic Immunity
The SDC Ultra™ system is intended for use in the electromagnetic environment specified below. The
user of the SDC Ultra™ system should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment--
Guidance
Portable and mobile RF communications
equipment should be used no closer
to any part of the SDC Ultra™ system,
including its cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended Separation Distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
d = 1.17√P
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.17√P 80 MHz to 800 MHz
d = 2.33√P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey (a), should be less that the
compliance level in each frequency range
(b).
Interference may occur in the vicinity
of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the SDC Ultra™ system is used exceeds the applicable RF
compliance level above, the SDC Ultra™ system should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the SDC Ultra™ unit.
(b) Over the frequency range 150 kHz to 80 Mhz, field strengths should be less than 3 V/m.
Содержание 0240050988
Страница 1: ...SDC Ultra 0240050988...
Страница 2: ......
Страница 4: ......
Страница 42: ...38 1 1 This shows the playback getting started 1 1 This shows the playback in progress...
Страница 43: ...39 1 1 This shows the playback is paused...
Страница 53: ...49 User Management Add Remove Users 1 1 Enable security for the system DICOM Settings 1 1 DICOM connection name...
Страница 55: ...51 SCU Setting 1 1 Service Class User SCU Settings 1 1 Shows all mandatory fields as defined by the PACS server...
Страница 73: ......