11
Warning
When used inside the patient environment (represented by an
area 1.5 m from the patient in all directions), the printer must
be properly connected to the isolation transformer prior to use.
Failure to do so may result in a higher leakage current for the
printer.
Note
See the “Technical Specifications” for approved printers and
transformers.
3. Connect the AC power.
• Connect the provided power cord to the AC inlet on the rear
console panel.
• Connect the other end to a hospital-grade power outlet.
Warning
Do not use extension cords or three-prong to two-prong adapt-
ers when connecting the AC power.
Warning
Check the power cord assembly periodically for damaged insu-
lation or connectors.